16
CD Indicators
3.2 Readout
The Readout menu is used to adapt the Indicator to environmental conditions. It contains five submenus:
Stable
,
Auto 0, Filter
,
Sleep, Lock
and
Exit.
Lock enables you to lock the settings. Review the settings available below
before proceeding.
Stability
The stability range specifies the weighing results and must be within a preset tolerance limit for a certain time to turn
the stability indicator ON. When a displayed weight changes beyond the allowable range, the stability indicator turns
OFF, indicating an unstable condition. Factory default setting is shown in bold type.
.
5 d
Smallest range: stability indicator is ON only
when displayed weight is stable within .5 divisions.
1 d
Normal range.
2 d
Medium range.
5 d
Largest range, stability indicator is ON even though displayed weight
changes slightly.
When the RS232 communication parameters are configured to print stable data only, the stability range also governs
data output. Displayed data will only be output if it is within the selected stability range.
Auto-0
(Auto-Zero)
Auto-0 minimizes the effects of temperature changes and shift on the zero reading. The Indicator maintains the zero
display until the threshold is exceeded. Factory default setting is shown in bold type.
OFF Turns Auto-Zero OFF.
.5 d Sets threshold to .5 divisions.
1 d Sets threshold to 1 division.
3 d Sets threshold to 3 divisions.
Filter
Filter compensates for vibration or excessive air currents. Default settings is shown in bold type.
-0- reduced stability, fastest stabilization time
-1- normal stability, normal stabilization time
-2- more stability, slow stabilization time.
-3- maximum stability, slowest stabilization time.
Sleep On/Off
When set ON, power will be removed from only the display in 5 minutes after last weighing. When set OFF, power
remains on all of the time to the Indicator circuitry and display. To turn power ON, press
Enter
.
Lock
Lock ON/OFF can only be changed when the hardware Lock Switch is set OFF/disabled (jumper on). A menu is
locked when the menu software lock is set ON and the physical Lock Switch is enabled. Lock (jumper removed), when
selected and turned on, locks all of the entries made under the Readout menu. In the locked condition, items may be
looked at in the menu but not changed. When set off, entries may be changed.
OFF
is the default setting.
NOTE
: When Lock Switch jumper on the bottom of the Indicator is in place, Lock is OFF and Readout menu
can be
changed. When Lock Switch jumper is removed, Lock is ON and Readout menu
cannot be
changed.