2.7 Initial Setup (Cont.)
2.7.3 Load Cell Capacity Information
Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the
indicator are compatible with the scale base. Below is a Load Cell Scale Capacity (lb or kg) table. Use this table to
determine the settings of the Indicator based on the capacity and resolution of the scale base.
LOAD CELL SCALE CAPACITIES (LB OR Kg)
Grad
1000d
1200d
1500d
2000d
2500d
3000d
4000d
5000d
6000d
7500d
10000d 20000d
0.001
-
-
-
-
-
-
-
5
-
-
10 20
0.002
-
-
-
-
-
-
-
10
-
-
20 40
0.005
5
-
- 10
-
-
-
20
25
30
50 100
0.005
5
6
7.5
10
12.5
15 20 25 30 37.5 50 100
0.01 10
-
-
20
25
30
40
50
60
75
100 200
0.02 20
-
30
40
50
60
-
100
120
150
200 400
0.05 50
60
75
100
-
150
200
250
300
-
500 1000
0.1 100
120
150
200
250
300
400
500
600
750
1000 2000
0.2 200
-
300
400
500
600
-
1000
-
1500
2000 -
0.5 500
600
750 1000
-
1500
2000
2500
3000
-
5000 10000
1
1000
1200
1500 2000
2500
3000
4000
5000
-
7500
10000 20000
2
2000
-
3000
-
5000
-
- 10000
-
-
20000
-
5
5000
-
7500 10000
-
- 20000
-
-
-
-
-
CD-11 Indicator
12
2.7 Initial Setup (Cont.)
2.7.4 Setup Menu
The CD-11 Indicator Setup Menu
must be entered the first time
the Indicator is used to set the scale base param-
eters to match the Indicator.
Do not attempt to calibrate the Indicator
before setting up the Setup Menu. All other
menus should be entered and set up the first time the Indicator is used.
Procedure
• With the Indicator ON, press and hold the
Zero/Menu
button until MENU is displayed. When you release
Zero/Menu
button, CAL is displayed when the CAL
jumper on the PC board is in place. When the CAL
jumper is removed, the Indicator will not permit cali-
bration. This jumper should be in place initially.
• Press
Print/Units
button, SETuP is displayed.
• Press
Zero/Menu
button, LFTOFF is displayed. legal
for trade selections are:
'ON' - LFT is ON
'OFF' - LFT is OFF.
• Press
Print/Units
button and select either ON or
OFF.
• Press
Zero/Menu
button, 0 2 is displayed. This is
the Zero 2%, 18% or 100% setting. 2% - zero opera-
tion range is - 2% to + 2%. 18% - zero operating
range is -2% to +18%, 100% - zero operation range
is -2% to +100%.
NOTE:
If LFT is ON, only 2% and
18% are available.
• Press
Print/Units
button, and select either 2%, 18% or
100%.
• Press
Zero/Menu
button, CAL Un kg is displayed.
This is the calibration unit setting. Selections are:
'lb' - calibration unit is lb
'kg' - calibration unit is kg.
• Press
Print/Units
button, and select either kg or lb.
• Press
Zero/Menu,
F xx is displayed
.
This is full
scale capacity selections. xx= value last set.
Selections are:
5, 10, 20, 25, 30, 40, 50, 60, 75, 100, 120, 200, 250,
300, 400, 500, 600, 750, 1000, 1500, 2000, 2500,
3000, 5000, 7500, 10000, 20000 (lb or kg).
• Press
Print/Units
button until desired capacity value
is reached.
• Press
Zero/Menu
button, Gd0.01 is displayed. This
is the graduation size. Refer to paragraph 2.7.3 Load
Cell Capacity Information table. For available selec-
tions, press
Print/Units
button until desired gradua-
tion value is reached.
• Press
Zero/Menu
button, CP 30 kg is displayed.
This is the full scale calibration point setting. The
range is from 20% to 100% Full scale capacity. Press
Print/Units
button until desired calibration value is
reached.
• Press
Zero/Menu
button to end this block, END is
displayed.
• Press
Zero/Menu
button, rEAD is displayed which is
the next menu or press
Print/Units
button to return to
Setup menu.
The Indicator is now matched up with the scale base and
the Indicator parameters may now be set and calibrated.
CD-11 Indicator