PAGE 11 OF 30
0908 IH-606
H-606
13
2.7 Initial Setup (Cont.)
2.7.5 Readout Menu
The Readout menu is used to adapt the Indicator to environmental conditions, set measuring units on/off, parts
counting, auto zero tracking, timer on/off and retain zero data. It contains 10 submenus. Review all of the settings
available before proceeding.
Procedure
To select any of the items in the Readout menu, proceed
as follows:
NOTE
: If you have entered from the
preceeding menu, disregard the first step.
• With the Indicator ON, press and hold the
Zero/Menu
button until MENU is displayed. When you release
the
Zero/Menu
button, CAL is displayed, then press
Print/Units
button, until rEAd is displayed.
• Press
Zero/Menu
button, rESETn is displayed. This
allows resetting the readout menu to factory defaults.
rESETn = no does not reset settings. rESETy= yes
will reset the entire readout menu as follows: AL Lo,
StAb 0.5, UnOff g, Un On kg, Un On Lb, AZt 0.5, Aot
Off, rZd Off.
• Press
Print/Units
button,and select N or Y.
AVERAGING LEVEL
Averaging level compensates for vibration or excessive
air currents on the scale base. During operation, the
indicator continually takes weight readings from the load
cell. Successive readings are then digitally processed to
achieve a stabilized display. Using this feature, you
specify how much processing you need.
HI and LO settings are available.
HI setting:
More processing, greater stability and slower
stabilization time.
LO setting:
Less processing, less stability and faster
stabilization time.
• Press
Zero/Menu
button, AL LO is displayed. This is
averaging level settings. Selections are:
'Lo' - Averaging level is low
'Hi' - Averaging level is high.
• Press
Print/Units
button,and select LO or HI.
STABILITY
The stability range specifies the weighing results and
must be within a preset tolerance limit for a certain time to
turn the stability indicator ON. When a displayed weight
changes beyond the allowable range, the stability indica-
tor turns OFF, indicating an unstable condition. Factory
default setting is shown in bold type.
.5d Smallest range: stability indicator is ON
only when displayed weight is within .5
divisions.
1d Normal
3d More stable course
CD-11 Indicator
14
2.7 Initial Setup (Cont.)
2.7.5 Readout Menu (Cont.)
Procedure
STABILITY (Cont.)
• Press
Zero/Menu
button, StAb1 is displayed.The
stability range specifies the weighing results and must
be within a preset tolerance limit for a certain time to
turn the stability indicator ON. When a displayed
weight changes beyond the allowable range, the stabil-
ity indicator turns OFF, indicating an unstable condition.
0.5 d smallest range, stability indicator is ON only
when displayed weight is stable within 0.5 divisions.
1 d- stable within 1 division. 3 d-largest range, stabil-
ity indicator is ON even though displayed weight
changes 3 divisions. Factory default setting is 1.
• Press
Print/Units
button,and select 0.5,
1
, or 3. Normal
1d stability is default/recommended.
UNITS SELECTION
• Press
Zero/Menu
button, Un OFF g is displayed. This
displays only when graduation is preset in grams.
• Press
Print/Units
button, and select ON or OFF. OFF
is the default setting.
• Press
Zero/Menu
button, Un ON lb is displayed. This
is unit pounds which can be turned ON or OFF. This
will be displayed when CAL UNIT kg was selected.
When lb was selected as calibration unit, kg will
display.
• Press
Print/Units
button,and select ON or OFF. ON
is the default setting.
• Press
Zero/Menu
button, Un (Pcs) OFF is displayed.
This is the Parts Counting function which can be turned
ON or OFF. Default setting is OFF.
• Press
Print/Units
button,and select ON or OFF.
AUTO ZERO
• Press
Zero/Menu
button, AZt 0.5 is displayed. This is
the Auto Zero Threshold setting. Auto Zero minimizes
the effects of temperature changes and small
disturbances on the zero reading. The Indicator
maintains the zero display until the threshold is ex-
ceeded. Settings are shown as follows:
0.5d Sets threshold to 0.5 divisions.
1d Sets threshold to 1 division.
3d Sets threshold to 3 divisions.
Factory default setting is 0.5d.
• Press
Print/Units
button,and select 0.5, 1 or 3.
Pcs
CD-11 Indicator