PAGE 23 OF 30
0908 IH-606
H-606
25
4.13 Printing Data
Printing data to an external computer or printer requires that the communications parameters in the Print menu, be
set first. Refer to paragraph 2.7.6 Print Menu for proper set up.
• To print data, press
Print/Units
button with a short press. The display acknowledges by momentarily blinking off.
NOTE
: If you hold this button down too long, the display will advance to another measuring unit.
4.12 RS232 Commands
All communication is accomplished using standard ASCII format. Characters shown in the following table are ac-
knowledged by the Indicator. Invalid command response "ES" error indicates the Indicator has not recognized the
command. Commands sent to the Indicator must be terminated with a carriage return (CR) or carriage return-line line
feed (CRLF). Data output by the Indicator is always terminated with a carriage return - line feed (CRLF).
4.12.1 Output Formats
Data output can be initiated in one of two ways; 1) By pressing
Print/Units
button, 2) Sending a print command (“P”)
from a computer or a PLC.
The output format is illustrated in the RS232 command table which follows.
Command
Character
Description
?
Print current mode: g, kg, lb
0S
Send weight (stable or unstable)
1S
Send stable weight only
AS
Auto send data when stable after motion
Z
Same as zero key. When scale stable, new zero established OK! response on PC screen.
When unstable, zero was not established, no response.
P
Print display data.
T
Same effect as pressing Tare Key. When scale stable, tare was established or cleared -
response OK. When unstable, "Cannot Tare" on display.
RS232 COMMAND TABLE)
4. OPERATION (Cont.)
CD-11 Indicator
26
5 CARE AND MAINTENANCE
To keep the Indicator operating properly, the housing should be kept clean and free from foreign material. If neces-
sary, a cloth dampened with a mild detergent may be used.
5.1 Troubleshooting
SYMPTOM
Unit will not turn on.
Cannot zero Indicator, or will not
zero when turned on.
Center of Zero display indicator
erratic or does not appear with no
load on scale base.
Cannot display weight in desired
weighing unit.
RS232 not working.
Unable to calibrate unit.
REMEDY
Check power cord connections.
Make sure adapter connector is
plugged all the way into the Indicator.
Check battery connector.
Check orientation of the batteries.
Replace batteries.
Check functions of membrane
switch.
Remove load on scale base to less
than entered zero %.
Change allowable zero % in ZERO
parameter of Setup menu.
Normal operation when this feature
is enabled.
Remove disturbances or reduce
motion.
Increase AZT level in readout menu.
Increase averaging level in readout
menu.
Enable desired unit in Readout
menu. See paragraph 2.7.5
Conversion to large (typically in g).
Verify communication parameters.
Check cable connections.
Check connections.
Set LCL to OFF in the LocSW menu,
and set Jump CAL on the circuit
board to short position. Refer to
paragraphs 2.3 and 2.7.7.
Use correct calibration mass.
PROBABLE CAUSE(S)
Adapter not plugged in or properly
connected.
Batteries dead or not properly
installed.
Membrane switch failure.
Load on scale base exceeds
allowable zero % entered in ZERO
parameter of Setup menu.
Retain Zero Data is enabled in
scale menu.
Scale base motion or disturbances
exceed center of zero criteria.
Desired unit not set to ON in
Readout menu.
RS232 communication parameters
set up incorrectly.
Improper or loose cable connec-
tions.
Scale base disconnected.
SETUP Lockout switch set to ON
and jump CAL on the circuit board
set to open position.
Incorrect value for calibration
mass.
CD-11 Indicator