PAGE 14 OF 30
0908 IH-606
H-606
15
AUTO POWER OFF
• Press
Zero/Menu
button, AOtOFF is displayed. This
is the Auto Off Timer. When set ON, the Indicator will
shut off automatically after 5 minutes has elapsed on
the condition that no button is pressed and the scale
base is stable during that period.
• Press
Print/Units
button,and select ON or OFF.
OFF is the default setting.
RETAIN ZERO DATA
• Press
Zero/Menu
button, Un rZdOFFis displayed.
This is Retain Zero Data which can be turned on or
off. When set On, the Indicator stores the current
zero point and restores it on the power-up.
• Press
Print/Units
button, and select ON or OFF.
OFF is the default setting.
• Press
Zero/Menu
button to end this block, END is
displayed.
• Press
Zero/Menu
button, Print is displayed which is
the next menu or press
Print/Units
button to go back
to Readout menu.
(If initial setup, go to the next paragraph. To exit from
the Setup, press
Prints/Units
button to skip to PRINT
then to LOCKSW, then QUIT. Press
Zero/Menu
button to go back to the weighing mode).
2.7 Initial Setup (Cont.)
2.7.5 Readout Menu (Cont.)
Procedure
CD-11 Indicator
16
2.7 Initial Scale Setup (Cont.)
2.7.6 Print Menu
The Print menu provides data communication settings which can be entered. It contains 6 submenus: Reset, Baud
rate, Parity Bit, Data Length, Stop Bits and End/Store.
Procedure
To select any of the items in the Print menu, proceed as
follows:
NOTE
: If you have entered from the preceeding
menu, disregard the first step.
• With the Indicator ON, press and hold the
Zero/Menu
button until MENU is displayed. When you release
the
Zero/Menu
button, CAL is displayed, then press
Print/Units
button, until Print is displayed.
• Press
Zero/Menu
button, rESEtn is displayed. This
allows resetting the Print menu to factory defaults.
rESETn = no does not reset settings. rESETy= yes
will reset the entire Print menu as follows:
Baud rate =2400, parity =none, data length=7,
stop bit=2.
• Press
Print/Units
button, and select N or Y.
• Press
Zero/Menu
button, bd2400 displayed.
• Press
Print/Units
button, and select desired baud
rate. Baud rate selections are: 1200, 2400, 4800 and
9600. 2400 is the default setting.
• Press
Zero/Menu
button, PAr NO is displayed. This
is the parity bit.
• Press
Print/Units
button, and select desired parity of
NO=none, Odd=odd, E=even. Default setting is none.
• Press
Zero/Menu
button, dAtA 7 is displayed. This
is the data length.
• Press
Print/Units
button, and select desired data
length of 7 or 8. Default setting is 7.
• Press
Zero/Menu
button, StOP 2 is displayed. This
is the stop bit.
• Press
Print/Units
button, and select desired stop bit of
1 or 2. Default setting is 2.
• Press
Zero/Menu
button to end this block, END is
displayed.
• Press
Zero/Menu
button, LOCSW is displayed which
is the next menu or press Print Units button to go back
to the Print menu.
(If initial setup, go to the next paragraph. To exit from
the Setup, press
Prints/Units
button to skip to PRINT
then to LOCKSW, then QUIT. Press
Zero/Menu
button to go back to the weighing mode).
CD-11 Indicator