Service Manual
Page 10 of 50
1.8. Sealing
After installation is completed, legal-for-trade applications require that the enclosure is sealed with a sticker.
Metrological relevant settings can
’
t be adjusted without breaking the sealing.
Attention:
Assuming the user is to use the (non-automatic) scales/balance in the legally regulated field, the user will
be responsible for notifying the appropriate calibration authorities of the repaired scales/balance, so that the
latter can take the appropriate measures (calibration/recalibration).
1. Sealing sticker
2. Sealing screw
3. Sealing cover