![OHAUS AV412 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/ohaus/av412/av412_instruction-manual_4150116038.webp)
Adventurer Pro
EN-36
3.9 Sealing Access to the Balance Settings
When used in conjunction with the Lockout menu, the balance may be sealed to prevent or detect
unauthorized changes to the balance settings. For legal for trade applications, the balance must
be sealed to prevent access to the metrological parameters.
To seal the balance, cover the Lock switch access hole at the rear of the balance. A paper seal, and
a wire seal may be installed as shown.
To regain access to the locked balance settings, break the seal and press the
recessed Lock switch momentarily while OHAUS is displayed during power up.
Sealing the Balance
6($/
Wire Seal Paper Seal
Rear of the Balance
3.10 Printing Data
Printing data to an external computer or printer requires that the communication parameters in the
RS232-1 and/or RS232-2 submenus be set to match external device communication parameters.
04/01/05 12:30 PM
// if GLP SET> Time is ON
Bal ID 1234567
// If GLP SET> Balance ID is ON
USER ID ABCDEFGHIJ
// If GLP SET> User ID is ON
PROJ ID 1234567890
// If GLP SET> Proj ID is ON
Name....................... // If GLP SET > Name is ON
// If a tare value is entered, these three items are printed if selected in the definitiion string.
0.0200kg G
// Gross ON - G, B or [space] as determined in Readout submenu.
0.0200kg T
// Tare ON
0.0000kg NET
// Net ON
// Otherwise, gross is printedif selected in the definition string.
0.0200kg G
// Gross ON - G, B or [space] as determined in Readout submenu.
Lock Switch