![OHAUS AV412 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/ohaus/av412/av412_instruction-manual_4150116028.webp)
Adventurer Pro
EN-26
NOTICE:
Linearity Calibration is required only if the linearity error exceeds tolerance in
specification table.
With the balance on, press and hold the
Menu-Cal
button
until MENU CALIBRATE is displayed, then release the button.
CALIBRATE (blinking).
Press the
Yes
button to enter calibration.
Note
: If the balance has InCal, press the
No
button to advance
to linearity calibration. CAL LINEARITY (blinking) is displayed.
Press the
Yes
button and follow the screen instructions. Use
the weight values shown on the primary display.
First the zero reading is taken. Next the display shows the first
calibration weight value. PUT WEIGHT (blinking) is displayed.
Place the specified calibration weight on the platform.
After a few seconds, place the second specified calibration
weight on the plattform. PUT WEIGHT (blinking) is displayed.
After a few seconds, the calibration weight is displayed.
LINEAR DONE is momentarily displayed when calibration is
completed.
After calibration, remove weight from platform. To leave the
menu, press
Exit
.
To abort calibration at any time, press the
Exit
button.
Linearity Calibration