APPENDIX B. SOFTWARE SERVICE & REPAIR TOOL INSTRUCTIONS
Adven
turer ™ Balances Service Manual 99 Ohaus Corporation www.ohaus.com
B.8
Diagnostics
1. Connect the balance to your computer.
2. Start the OHAUS Software Service Tool.
3.
Click on the tab labeled “
Diagnostics.
”
Figure B-12. Diagnostics Tab.
4. To test communications, click the Print button, or enter an RS232 Command String (from
Table 1-6) and USB Input (1-7) and click the
Send
button.
Figure B-13. Diagnostics Command Testing.
5.
The scale’s response will be shown in the box on the right.