APPENDIX B. SOFTWARE SERVICE & REPAIR TOOL INSTRUCTIONS
Adven
turer ™ Balances Service Manual 97 Ohaus Corporation www.ohaus.com
15. Configure the Com Port Configuration (B.2) and click 'Next'.
16. Enter the IDNR number (xxx.xx.xx.xxxx) from the label on the back of the balance.
17. Browse to the folder where the balance image file (eif) is store.
18. Click the
Start
button.
19. The software will indicate the download progress. When complete, disconnect the power
from the back of the balance, then re-connect.
20. Perform Service Calibrations (see Appendix B). Test the balance.
B.6
Install New Main Printed Circuit Board (PCBA).
1. Follow steps in Sections 3.4 to replace the PCBA. Then connect the balance to your
computer.
2. Start the OHAUS Service and Repair Tool Software.
3. Click on the tab labeled
Replace Main PCBA.
Figure B-9. Replace Main PCBA Tab.
Figure B-10. Function Selection Replace Main PCB.