Analyzer and Bench Meter
EN-23
8. Menu
Touch the
Menu
icon
to enter the meter/analyzer's setup menu.
8.1
Setup menu
Enter this menu to set the meter/analyzer's parameters. The default settings are in bold.
Table 8-1 Setup menu
Menu
Sub-Menu
Sub-Menu
Options
Options
Setup
Mode
Read type
Auto-Stop
, Continuous, Timed (a-
AB41PH)
Stability
SmartStability
, Fast, Medium, Slow (a-
AB41PH)
Average
Auto Smart
, OFF
pH
(a-AB33M1,
a-AB33PH,
a-AB41PH)
Resolution
0.1,
0.01
, 0.001 (a-AB41PH)
Buffer group
USA
, JJG, DIN, EU, Merck
Conductivity
(a-AB33M1,
a-AB33EC)
Cell K
0.01~10.0 (the default is
0.1
(a-AB33EC)
and
1.75
(a-AB33M1))
Std.Solution
10 μS/cm,
84 μS/cm
(a-AB33EC), 146.5
μS/cm, 500 μS/cm,
1413
μS/cm
(a-
AB33M1) and 12.88 mS/cm.
Ref.Temp.
20°C(68°F),
25°C(77°F)
Temp.Coeff.
0.0~10.0%/
℃
(the default is
2.0%/
℃
)
TDS Factor
0.01~5.0 (the default is
0.50
)
Temperature
Temp.unit
°C
, °F
Enter
temp.value
5.0~60.0
℃
(41.0~140
℉
)
(the default is
25.0
℃
/77.0
℉
)
Temp.cal
(a-AB41PH)
0~100
℃
(32~212
℉
)
Read type
Set the endpoint mode.
Auto-Stop
= the meter/analyzer determines when the reading is stable and
reaches the endpoint automatically.
Continuous
= the meter/analyzer keeps measuring and users manually end the
measurement.
Timed (a-AB41PH) = the meter/analyzer keeps measuring during the time period
users set.
Note
: for detailed information about the three endpoint modes, please refer to
Measurement modes
section.
Stability
Set measurement accuracy.
SmartStability
= the meter/analyzer determines stabilization time automatically.