CR Series Scales EN-3
3. OPERATION
3.1 Two Button Keypad
:
Pressing this button turns on the scale. This same button operates
the zero feature when the scale is on. Press and hold this button for three
seconds to turn the scale off.
:
Press this button briefly to change the weighing unit. Press and
hold this button to begin the calibration process.
3.2 Calibration
For best results calibrate the scale at regular intervals. This is especially important
if the scale is in use for prolonged periods.
Calibration weights are not provided with the scale
Press and hold the
button to start the
calibration process. The display will show
CAL
.
The calibration process can be aborted by turning the
scale off.
Press the
button to capture 0. The display
shows
-C-
while the scale stores the zero load
signal.
The display will show
C
xxx where xxx is the
calibration weight in grams.
Place the appropriate calibration weight on the
platform.
Press the
button.
The display shows
-C-
while the scale stores the
calibration point signal.
After calibration, the display returns to the normal
weighing mode.
The message
CAL E
will appear if the calibration steps are not
followed or the wrong weight was used.
www.
GlobalTestSupply
.com
Find Quality Products Online at: