Internal Component Description 

 

Equalization Valve A & B – 

The Equalization Valves allow gas to pass between the beds at the end of each 
pressurization cycle in order to increase the efficiency of the machine by 
reducing the amount air needed to produce a unit of Oxygen. 

 

Feed Valve A & B – 

The Feed Valves allow Air to enter the machine and pressurize the beds 
beginning the air separation/oxygen generation process. Only one of these 
valves is open at any point in time. 

 

Waste Valve A & B – 

The Waste Valves allow the Nitrogen gas that has been trapped during the 
previous adsorption cycle to be discharged back to the air.  As with the Feed 
Valves only one of these valves will be open at a time.  They operate in 
conjunction with the opposite side Feed Valve.  For example, whenever the Feed 
Valve on the A side of the machine is open, the Waste Valve on the B side will 
also be open.  The reverse is also true. 

 

Drain Valve – 

The Drain Valve allows condensate and waste material trapped in the filter to be 
discharged automatically at fixed intervals by the PLC or at any time by 
depressing the Manual Drain button. 
 

Pressure Switch – 

The Pressure Switch senses the Oxygen Pressure in the Oxygen Storage Tank 
and will turn the Oxygen Generator On and Off automatically when it is in the 
‘Automatic’ mode.  The Pressure Switch comes set from the factory with a ‘Turn 
Off’ point at 60 psig and a ‘Turn On’ point at 45 psig.  Both the dead band 
(difference between the Turn Off and Turn On points) and the points themselves 
can be adjusted in the field or by our factory technicians prior to shipment.  
Please call the factory for assistance in this area if you want to make changes. 

 

Filter – 

The filter keeps dust, dirt, and moisture from entering the sieve beds and 
damaging the sieve. 

 

 

Содержание OG-75

Страница 1: ...814 Wurlitzer Drive North Tonawanda New York 14120 Telephone 716 564 5165 Fax 716 564 5173 Toll Free 800 414 6474 E mail ogsimail ogsi com Web Site http www ogsi com Model OG 75 thru OG 750 Oxygen Ge...

Страница 2: ...onnection Diagram 14 Setup Instructions 15 16 Safety Precautions 17 Start Up Procedures 18 19 Shut Down Procedures 20 Troubleshooting Guide 21 Detailed Warning Description 22 Component drawings descri...

Страница 3: ...2003 OGSI All rights reserved OGSI is a registered trademark This publication may not be reproduced in part or whole without written permission of OGSI...

Страница 4: ...unit to the same degree that it was received from OGSI to ensure its undamaged arrival at OGSI If any damage is discovered during the initial visual inspection call the Freight Company immediately Th...

Страница 5: ...which is Greenwich Mean Time minus 5 hours GMT 5 You may reach OGSI personnel through the following means By Telephone from within the United States two numbers may be used 800 414 6474 our toll free...

Страница 6: ...ned from OGSI prior to return shipment of equipment These warranties shall also become null void and not binding on OGSI if a defect or malfunction occurs in the product or any part of the product as...

Страница 7: ...combustible materials Precautions should be taken to avoid a fire in the area of the generator Smoking should not be permitted in the area where the generator is located All oxygen connections and hos...

Страница 8: ...d not be removed except for active use Remember to replace removable caps when not in use Do not place cylinders in a position where they might become part of an electric circuit When electric welding...

Страница 9: ...f the beds the sieve in that bed holds the Nitrogen to it and allows the Oxygen to flow through it and out to the surge tank as product gas Eventually the sieve becomes saturated with Nitrogen When th...

Страница 10: ...183 775 351 OG 250 26 x 32 x 72 66 x 81 x 183 1100 499 OG 375 30 x 36 x 84 76 x 91 x 213 1350 612 OG 500 36 x 48 x 83 91 x 122 x 211 1800 816 OG 650 48 x 55 x 83 122 x 140 x 211 2400 1625 OG 750 38 x...

Страница 11: ...ee of toxic gases and high concentrations of hydrocarbons especially carbon monoxide Additionally to the degree possible humid oily areas should be avoided as installation sites Ambient Air Temperatur...

Страница 12: ...eep it from flowing into the generator Finally as already mentioned in a few areas the filter element inside the Oxygen Generator needs to be replaced every six months Electrical Power On U S models t...

Страница 13: ...ted G 22 O I S G 22 G I OGS O I S G 22 G Occasionally a user may choose to maintain a liquid or cylinder backup and keep it connected to his system for critical applications This is a precaution that...

Страница 14: ...making sure no empty voids are left in the beds The sieve can be harmed by exposure to air and caution must be taken to keep the sieve in the oxygen generator the container sealed as much as possible...

Страница 15: ...Effective April 2006 12 Sieve Re packing Diagram...

Страница 16: ...FRONT OF THE ENCLOSURE DOOR 3 NEXT TAKE 6 AND PUSH ONTO 1 FROM THE BACKSIDE OF THE ENCLOSURE DOOR 4 NOW HAND TIGHTEN 5 ONTO 1 FROM THE BACKSIDE OF THE ENCLOSURE DOOR 5 AFTER THIS PUSH 3 INTO 1 FROM TH...

Страница 17: ...TCH ASSEMBLY INSTRUCTIONS 1 TAKE 1 AND PUSH THROUGH 4 2 NOW TAKE 1 4 AND PUSH THROUGH THE FRONT OF THE ENCLOSURE DOOR 3 PUSH 2 ONTO 1 AND TURN IT TO LOCK THEM TOGETHER 4 NOW MAKING SURE THE ASSEMBLED...

Страница 18: ...or a male pipe thread depending on which model you have Connections to these are readily available worldwide through the Parker distribution network or from OGSI 3 Unroll and Direct the Filter Drain C...

Страница 19: ...before this Oxygen Hose and Ball Valve are not included with the generator because the connections used from location to location can vary greatly We can provide these Ball Valves and Hose Assemblies...

Страница 20: ...acility 2 Ensure that the oxygen outlet stream is not directed toward anyone s clothing Oxygen will embed itself in the material and one spark or hot ash from a cigarette could ignite the clothing vig...

Страница 21: ...lown out the drain valve on the bottom of the cabinet If after 5 seconds air is not blowing out the filter drain then manually depress the filter drain push button on the lower left hand side of the c...

Страница 22: ...cation This valve although not included in our standard accessory package should be in place to ensure a proper shut down and avoid the possibility of wasting oxygen through an improperly closed torch...

Страница 23: ...tch to Off This should really only be done if the storage tank is already pressurized to at least 40 PSIG 280 kPa OR 2B Leave the Control Switch in the AUTO Position and Allow the Generator to Cycle U...

Страница 24: ...high purity The following troubling shooting checks are for machines equipped with the optional Touchscreen Control system only System fails to start 1 If in manual mode press the additional start pu...

Страница 25: ...set to 50 psig higher than the maximum setting for normal system operation High pressure transducer replacement will be necessary if this fault occurs Low Oxygen Pressure This may be a result of a lea...

Страница 26: ...Effective April 2006 23 Process Flow Schematic...

Страница 27: ...Effective April 2006 24 Process Flow Schematic Touchscreen Option...

Страница 28: ...is directed into one of the two sieve beds As the air enters the bed the nitrogen is adsorbed by the sieve and the oxygen passes through as product gas Each bed produces Oxygen until the sieve in that...

Страница 29: ...Effective April 2006 26 External Components Drawing 22 G G 22...

Страница 30: ...of whether or not oxygen is being drawn from the storage tank The light inside the switch will always be on as well The advantage to operating in this mode is that oxygen should be available at a slig...

Страница 31: ...ort circuiting either a fuse or circuit breaker depending on which model you have Generators shipped outside North America will have an electrical plug suitable for use in the location to which it is...

Страница 32: ...you should attach to it It can be used to set the regulated Oxygen pressure out of the storage tank at any level up to 60 PSIG 420 kPa Please keep in mind that our specification for the maximum oxyge...

Страница 33: ...Effective April 2006 30 Internal Components Drawing SMC SMC Direct 05 LOGIC Koyo SMC...

Страница 34: ...henever the Feed Valve on the A side of the machine is open the Waste Valve on the B side will also be open The reverse is also true Drain Valve The Drain Valve allows condensate and waste material tr...

Страница 35: ...he hours meter increments time while the unit is running It provides an indication as to when service intervals are due Noise Muffler The noise muffler is used to silence the exhaust noise that occurs...

Страница 36: ...quest The next screen is the Control Screen This screen contains the modes of operation and indicator buttons For the machine to start cycling one of the Mode of Operation Buttons must be selected eit...

Страница 37: ...een contains a graph of the 4 gauges over time Every five minutes a reading is taken The legend on the bottom of the screen contains the coordinating colors of the meters The Alarm Screen contains 7 t...

Страница 38: ...ow Area If you want SPANISH as your default language touch anywhere in the Yellow Area Once a language has been chosen press this button to advance to the next screen If you want ENGLISH as your defau...

Страница 39: ...apping the password area titled Enter Numeric Entry When this appears enter your numeric password given to you by OGSI Note The user will not be allowed to change their default password so keep it sec...

Страница 40: ...TATUS represents the Generator having proper supplied power The Indicator Button CONTINUOUS MODE represents if CONTINUOUS is selected in the main control If any other option is selected on the MAIN CO...

Страница 41: ...rator is in STANDBY MODE meaning the Pressure switch has gone active meaning the machine is NOT Cycling but the Generator is still in Automatic Mode The Indicator Button CYCLING represents if the gene...

Страница 42: ...ic Mode Simply touch any where in the green area and a numeric keypad will appear to enter the low point Only 2 digits can be placed into the keypad Some limitations are in the top right corner of the...

Страница 43: ...creasing it one hour accordingly You may adjust the REALTIME CLOCK by decreasing it one hour accordingly If you want to go back to the Previous Screen Control Screen all you have to do is touch the Bl...

Страница 44: ...e except for the normal operating variance that is between 0 and 20 psi This Meter OXYGEN PRESSURE measures the Oxygen Pressure psig in your back pressure regulator The same details hold true for this...

Страница 45: ...clear the readings and if the generator is not cycling you can stop the graph from reading Using the CLR and Stop Buttons Also you can scroll through the graph using the 4 buttons in the upper left co...

Страница 46: ...is above 80 psi This particular alarm is triggered when the Oxygen Pressure is below 40 psi This particular alarm is triggered when the Oxygen Pressure is above 65 psi This particular alarm is trigger...

Страница 47: ...ril 2006 44 If you want to go back to the Previous Screen Pressure Screen all you have to do is touch the Blue Area If you want to advance to the Next Screen Final Screen all you have to do is touch t...

Страница 48: ...GE purity of your OGSI Generator This is the date that your machine was shipped General Information If you want to go back to the Previous Screen Alarm Screen all you have to do is touch the Blue Area...

Страница 49: ...Effective April 2006 46 Manifold Drawing...

Страница 50: ...Oxygen under high pressure is present The use of safety glasses and hearing protection is required Slowly disconnect the steel whip from the cylinder valve using a 1 1 8 wrench Be aware of the ventin...

Страница 51: ...long runs condensation inside the pipes can be a big problem The solution is to have traps and drains opened regularly automatically if practical to keep the water from reaching your Oxygen Generator...

Страница 52: ...to pull it out of the fitting the collar of the Prestolok fitting is probably not being pushed back into the fitting hard enough The tube should come out easily once it is released from the fitting 7...

Отзывы: