OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
29
4.
Regulator and Manifold Maintenance:
a.
Replace both o-rings on the puncture pin (#143-02-13 and #143-02-
14) at least once a year.
b.
Regulator loses pressure or steadily increases in pressure – Replace
the seat assembly and diaphragm.
Concoa Repair Kit – Part #143-07. Normally the Concoa regulator is sup
-
plied with the 175 mL models
The Large High Pressure Concoa Regulator (805-1140) and the smaller
Concoa Regulator (805-1179) both take the same Regulator Repair Kit
Always replace (order separately) the rubber diaphragm, #143-00-1,
which is not supplied with the #143-07 kit.
200-psi Gauge
(#143-01)
#142-10 - CO
2
Pressuring Assembly
Safety Bleeder Valve
(#143-06)
Regulator (#143-00)
CO
2
Puncture Head
Assembly (#143-02-10)
CO
2
Cartridge
(#143-02-05)
Barrel for CO
2
Cartridge (#143-03)
CO
2
Puncture Head
Assembly (#143-02-10)
CO
2
Puncture Head
Assembly (#143-02-10)
Puncture Pin Holder
Assembly (#143-02-11)
O-ring
(#143-02-14)
O-ring
(#143-02-13)
Diaphragm
(#143-00-1)
Repair Kit
(#143-00-1)