background image

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

18

Sample Data

Graph 3

Graph No. 3 shows interesting micro views of the behavior of the different 

fluids up to 30 minutes into the test.

All fluids show a positive slope confirming an immediate interaction between 

the clay and the fluid bases.

The potassium acetate fluid initially showed the highest rate of expansion 

of the shale and this state continued up to around 28 minutes where the 

potassium acetate began acting as a better shale stabilizer base than the 

fresh water.

During the 15 to 20 minute time period ALL of the samples exhibited 

increased hydration as exhibited by the sudden increase in the slope of all of 

the curves.

Содержание 150-80

Страница 1: ...r 150 80 115V 150 80 1 230V Software Version 3 09 0 Instruction Manual Updated 6 28 2018 Ver 6 0 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 320 7300 Fax 713 880...

Страница 2: ...Software 6 Zeroing the Channels 8 Calibration 9 Starting a Test 10 Setup 12 Analog Input 13 Operation 14 Sample Data 15 Graph 1 16 Graph 2 17 Graph 3 18 Graph 4 19 Compactor 20 Assembly 20 Making Wafe...

Страница 3: ...riable Differential Transducer LVDT measures the expansion of the wafer in the vertical direction accuracy to 0 1 and this information is then stored as a function of time via the data acquisition sys...

Страница 4: ...up 150 80 036 Cap for Wafer Tube 150 80 064 LVDT 150 80 094 Cable LVDT to Swell Meter 150 83 Stirring Hot Plate 120 Volt Only 150 84 Stirring Hot Plate 230 Volt Only 152 37 AC Power Cord 3 Conductor 1...

Страница 5: ...ge 26 Adjust the height of the head assembly such that there is a small gap approximately 5 mm between the bottom of the nut on the spindle and the body of the LVDT without a wafer inserted into the c...

Страница 6: ...s located on the back of the Swell Meter are in the OFF position and make the necessary electrical connections in accordance to local codes Ensure that the unit is grounded 7 Make sure the computer is...

Страница 7: ...or later choose Swell Meter V2 If you are not sure look at the cable connecting the computer to the Swell Meter If it is a flat wide ribbon cable select Swell Meter If it is a standard USB cable sele...

Страница 8: ...r has been configured to run eight cells instead of just four you will need to change the Number of Cells Available option on the Setup screen to 8 This will enable to software to run all eight cells...

Страница 9: ...swell during the test 1 Click the Zero Start Stop button 2 Repeat the following steps for each channel to be tested a Enter a test name and any comments b Place two screens into a fully assembled cup...

Страница 10: ...ing on the bottom of the cup 2 In the Swell Meter software select Calibration from the Options menu 3 Choose the cell you want to calibrate and click the Calibrate Cell button Calibration Block 150 80...

Страница 11: ...Place a screen on both sides of the compacted wafer and insert it into the wafer tube with the transfer stand on top of it 4 Place the LVDT spindle on top of the transfer stand 5 Click the Zero Start...

Страница 12: ...that was recorded when the channel was zeroed This value is subtracted from the current raw distance reading to calculate the corrected distance inches mm Init Samp This is the initial distance of th...

Страница 13: ...alue in this field is set to 0 0010 a data point will be recorded whenever the corrected distance changes by more than 0 001 Units are determined by the Distance Unit field below Swell To record a dat...

Страница 14: ...rease in temperature To test the transducer move it up and down and observe the graph The distance line should move accordingly If the graph does not react as expected the communication between the Sw...

Страница 15: ...in the cap 7 Turn on left knob the heated magnetic stirrer To heat the fluid sample turn the right knob and read the temperature in the Temp field in the software The maximum temperature of the hot pl...

Страница 16: ...erred to as Rev Dust The wafers were compressed at 6 000 PSI for 30 minutes The results and interpretation of the graphs using micro and macro temporal criteria clearly explain the following phenomena...

Страница 17: ...ol acted as the better shale stabilizer exceeding all other fluids The amount of expansion stayed in a straight line of zero slope indicating a very positive balanced activity resulting in an ideal co...

Страница 18: ...x 713 880 9886 www ofite com 17 Sample Data Graph 2 Graph No 2 shows results obtained over an 8 hour time period and shows in greater detail the changes that occurred between the base fluids prior to...

Страница 19: ...confirming an immediate interaction between the clay and the fluid bases The potassium acetate fluid initially showed the highest rate of expansion of the shale and this state continued up to around...

Страница 20: ...te interaction between the clay and the fluid The fresh water initially showed the least amount of hydration of the four bases The yellow curve of potassium acetate had the highest rate of expansion a...

Страница 21: ...ed vessel capable of very high pressures extreme care should always be taken by the operator and anybody that is nearby The Compactor can make two wafers at a time Since the Compactor does not require...

Страница 22: ...rt the assembly and pour the sample to be tested into the small opening of the cell body The amount should not be less than 10 grams and not more than 20 grams 3 Insert the thin 14 mm spacer in the ce...

Страница 23: ...umping the hand pump handle will apply pressure to the Compactor The pedestal base and the cell assembly will begin rising in the Compactor Observe the pressure gauge on the front of the Compactor and...

Страница 24: ...and the spacer from the opposite end of the cell body 7 Drop the tall 20 mm spacer and position it in place where the thin spacer had been Re position the piston as before 8 Replace the cell assembly...

Страница 25: ...crest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 24 Drawings Cup Assembly Brass Weight Optional Transfer Stand Cap for Wafer Tube Bottom Plate Teflon Washer Wafer Tube Mag...

Страница 26: ...OFITE 11302 Steeplecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 25 Drawings Head Assembly...

Страница 27: ...OFITE 11302 Steeplecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 26 Drawings Main Assembly Cup Assembly Magnetic Stirrer Head Assembly...

Страница 28: ...OFITE 11302 Steeplecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 27 Drawings Wafer Die Specimen Removal Specimen Creation...

Страница 29: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Отзывы: