OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
28
d. Install a new contact arm and make adjustments as necessary to
obtain consistent contact with the resistor.
e. Loosen the three adjustment screws on the underside of the
potentiometer assembly and rotate the spring adjuster until the spring
rests at a relaxed state.
f. Ensure that the contact arm aligns with the contact strip and tighten
the three set screws.
g. Rotate the center shaft to ensure that the spring does not bind or rub
the potentiometer housing.
h. Replace the shaft-bearing retainer and calibrate the potentiometer.
Shaft-Bearing
Retainer
Contact Arm
Resistor
Calibration Spring
Set Screws