24
13. The back pressure regulator may contain some residual debris that must
be removed. To clean the device, apply pressure to force out any foreign
material:
a. If the back pressure regulator is still connected to the test cell,
disconnect it and remove the tube and quick-connect hose.
b. Connect the unit to the Back Flow Pressure via the quick-connect
hose. Attach the hose to the same port that was previously connected
to the valve stem on the test cell.
c. Close the Back Flow Pressure bleed valve on the right-hand side of
the control panel.
d. Gradually increase the Back Flow Pressure until the entire contents of
the back pressure regulator has been forced out.
Be sure to direct the flow of debris away from people and
equipment.
To Back Flow
Pressure Port
14.
Remove the flasks, beaker, and tubing from the inside of the unit cabinet
and thoroughly clean them with soap and water.