4.3.1 Transmitter
Powering
Voltage:
+24
VDC
±5
%
Current:
max.
100
mA
Fuse protection:
200 mA, self-resetting
Powering complies the requirements of a grounded
low voltage (SELV-E according to EN61010).
4.3.2 Measuring
Technique
Measuring ranges:
Sensor dependent (max. 1.5·10
3
…5·10
-9
mbar)
Measuring
rate:
50
s
-1
Display
rate:
4
s
-1
Units of measuring:
mbar, Pa, Torr
4.3.3 Switching
Function
Type of contact:
Changeover contact, potential-free
Load (resistive):
Switching current: max. 1 A
Switching voltage: max. 30 V AC / 60 V DC
Service
life:
Mechanical:
10
7
actuations
Electrical: 10
4
actuations at maximum load
Connection:
Plug-in terminals, Screw terminals
4.3.4 Analog
Output
Number:
1
Voltage range:
0…10
V DC
Deviation from the displayed
Value
:
±
0,1
%
Internal resistance:
100
Ohm
Relation between
voltage and pressure:
Sensor dependent
Connection:
Plug-in terminals, Screw terminals
10