PERIODIC MAINTENANCE
Customer Name
Street
City
State
Zip
Serial No.
MINI6600 IMAGING SYSTEM
FSR No.
Customer Signature
Service Engineer Signature
Date
Date
TEST EQUIPMENT USED
TYPE
MODEL
CAL DUE DATE
Employee #
Hard Copy Camera
VCR
Thermal Printer
S/N _______________
SYSTEM CONFIGURATION
Semi-annual PM
Annual PM
Remove Power and Covers
Elapsed Time Indicator
Inspect Static Drag Wire
Inspect the Articulating Arm
Inspect the Rotation Brake
Inspect the Orbital Brake
Verify System Steering and Brake Operation
Inspect/Clean the Workstation Air Filter
Inspect the AC Power Cable
Inspect the Cables, Connectors and PCBs
Measure Ground Continuity
Power Up the System
System Virus Inspection
Measure Line Voltage Regulation
Inspect/Adjust Workstation Power Supply Voltages
Inspect/Adjust Mini C-Arm Power Supply Voltages
Inspect the Service Outlet Strip
PERIODIC MAINTENANCE PROCEDURES
COMMENTS
Line Voltage Regulation = ______%
White - Customer
Yellow - Factory
Pink - Regional Service Center
Green - Service Engineer
Page 1
12-040-9/96
1. GENERAL INSTRUCTIONS
2. SAFETY
3. COMPLETING THE PM FORM
4. BEGIN PERIODIC MAINTENANCE
Semi-annual PM: Perform 1, 2, 3, 4, 6, 7
Annual PM Perform: 1, 2, 3, 4, 5, 6, 7
S/N _______________
Time of Day:
Elapsed Time Indicator : __________
WORKSTATION VOLTAGES
MINI C-ARM VOLTAGES
+24 VDC (+/- 1.0) ________ VDC
+5 VDC (+/- 0.1) ________ VDC
-5 VDC (+/- 0.75) ________ VDC
+12 VDC (+/- 1.0) ________ VDC
-12 VDC (+/- 1.0) ________ VDC
+5 VDC (+/- 0.10) ________ VDC
+12 VDC (+/- 0.36) ________ VDC
-12 VDC (+/- 0.36) ________ VDC
+24 VDC (+2.4,-1.2) ________ VDC
-10 VDC (+/- 0.2)________ VDC
Record Voltages Measured Below
Record Voltages Measured Below
Record Voltages Measured Below
115 VAC (+/- 5.0) ________ VAC
230 VAC (+/- 10.0) ________ VAC
S/N
4" I.I.
6" I.I.
WORKSTATION POWER STRIP
S/N _______________
COMMENTS
: