MINI6600
TM
Mobile Digital C-Arm Service Manual - Periodic Maintenance
Verify the System Controls
WARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
1. Connect the footswitch and the handswitch (optional) to the system. Confirm the operation of each
switch.
2. Place the Beam Alignment Tool on the Image Intensifier.
3. Verify the CAMERA rotation operation.
4. Verify the camera REVERSAL operation.
5. Verify the NEGATE operation.
6. Verify the MANUAL BRIGHTNESS and CONTRAST operations.
7. Verify the AUTO BRIGHTNESS AND CONTRAST operations.
8. Verify the MANUAL MODE kV/
µ
A operation.
9. Verify the AUTO NORMAL MODE operation.
10. Verify AUTO OSTEO/PED MODE operation.
11. Verify the image SAVE operation. Not applicable if system is equipped with 4 image storage capability.
12. Verify the VCR RECORD and the VCR PLAY operations (if applicable).
13. Verify the ONE-SHOT MODE.
a. System will fluoro until the video stable condition is achieved and then terminate the exposure.
14. Verify the ENHANCED MODE.
a. System will fluoro until the video stable condition is achieved and then terminates the exposure as
well as enhances edge.