MINI6600
TM
Mobile Digital C-Arm Service Manual - Periodic Maintenance
Begin Periodic Maintenance
Remove the Power And Covers
WARNING:
Electrical circuits inside the equipment use voltages that are capable of causing serious injury
or death from electrical shock. Use appropriate precautions.
1. If the system has power applied, position the System keyswitch/pushbutton to the OFF position. Verify
that the System’s AC power cord is not plugged into the wall outlet.
2. Remove the System covers. Refer to the System Overview section in the MINI6600 Digital Mobile C-
Arm Service Manual (00-878315) for cover removal procedures.
Elapsed Time Indicator
Read the elapsed time meter on the System and enter the number on the PM Form.
Inspect the Static Drag Wire
1. Verify that the static drag wire underneath the System is touching the floor.
2. Verify the static drag wire is securely attached.
Inspect the Articulating Arm and Brake
Verify that the articulating arm mechanics move freely and smoothly throughout the entire range of
motion.