Inspect/Clean the System Air Filter
Inspect and clean the System air filter attached to the rear panel.
Inspect the AC Power Cable
Inspect the System AC power cable for damage and wear.
Inspect the Cables, Connectors and PCB’s
1. Inspect all the cables for damage. Look for signs of abrasion to the cable jacket. If abrasions are de-
tected, replace the cable.
2. Inspect all connectors. Ensure that all connectors are seated securely.
3. Ensure all PCB’s are seated securely.
Measure the Ground Continuity
WARNING:
Do not perform this continuity test with the AC plug plugged into a power receptacle.
Verify the system power cord is disconnected from the AC receptacle.
From
To
Spec
System Equipotentiality Stud
(Underneath the bottom right side)
AC Power Plug Ground Pin
< 0.15
Ω
TABLE 1