COMPACT 7600 Periodic Maintenance, 00-878852-01

5

7. While taking x-rays in the Auto Fluoro mode press the slot collimator open and close buttons

 and verify that the collimator opens and closes.

8. While taking x-rays in the Auto Fluoro mode press the slot collimator rotate buttons

 and verify that the collimator rotates in both directions.

9. While taking x-rays in the Auto Fluoro mode press the iris collimator open and close buttons

 and verify that the iris opens and closes.

10.Press the Magnification button 

 repeatedly and verify that the LED turns on and off. With

Magnification active (LED on) verify that the iris collimator sizes down and the image on the
monitor changes to the small field mode when x-rays are taken.

11.Press the Image Reversal button 

 repeatedly and verify that the LED turns on and off.

With Image Reversal active (LED on) verify the image reveres horizontally when x-rays are
taken.

12.Press the Image Rotation buttons 

 

 and verify that the image on the monitor rotates

in both directions. Press both keys simultaneously and verify that the image rotates to the normal
orientation and automatically stops.

13.Press the Thorax Surgery button 

 repeatedly and verify that the LED turns on and off.

With Thorax Surgery active (LED on) take an x-ray and verify that a live image appears on the
monitor.

14.Press the Hand Surgery button 

 repeatedly and verify that the LED turns on and off. With

Hand Surgery active (LED on) take an x-ray and verify that the field size automatically decreases
because the iris collimator closes partially.

15.Place lead aprons on the image intensifier to attenuate the beam. Press the Hip Surgery button

 repeatedly and verify that the LED turns on and off. With Hip Surgery active (LED on)

take an x-ray and verify that the mA increases to a maximum of 6mA.

16.Obtain an image on the monitor and press the Save button 

. Verify that the image is saved

by noting the image storage number which will appear on the monitor.

17.Save 4 distinct images in memory using the Save button 

. Press the Image Recall button

 repeatedly and verify that the saved images are displayed in sequence on the monitor.

Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800 

An Independent C-Arm Service Provider.  This PDF provided for research / historical purposes only.

Содержание COMPACT 7600

Страница 1: ...Maintenance Procedure 00 878852 01 Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research his...

Страница 2: ...ne who is not specifically trained or authorized by OEC Medical Systems Inc to work on the COMPACT 7600 System The contents of this procedure are believed accurate at the time of publication However c...

Страница 3: ...996 Initial Engineering Release Rev 2 01 Aug 1996 Added fast stop button operational check Rev A 01 Aug 1996 Manufacturing Release Used and Refurbished C Arms are Available from www SharpMedical com C...

Страница 4: ...riodic Maintenance 00 878852 01 Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research histori...

Страница 5: ...4 5 Measure Line Voltage Regulation 3 4 6 Perform System Functional Test 4 5 Generator Accuracy Verification Calibration 6 5 1 Verify X Ray Calibration 6 5 2 Perform X Ray Calibration 6 6 Radiographic...

Страница 6: ...riodic Maintenance 00 878852 01 Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research histori...

Страница 7: ...Serious injury and property damage can result from incorrectly performed service procedures Observe all operating and safety procedures contained within this manual CAUTION The procedures in this sec...

Страница 8: ...1 Mechanical Checks In the steps below check for ease of movement without excessive play 1 Check the Locking Pedals of the system for proper operation Lock the system in place by pressing the locking...

Страница 9: ...and has a 9 pin connector on the end of the bundle The 9 pin connector is designated as P2 Measure the voltages listed below and record them on the PM Form GND To Adjust Voltage Tolerance P2 2 P2 1 R3...

Страница 10: ...button and verify that the system produces pulsed x rays and that an image appears on the monitor 4 Press the Snapshot button repeatedly and verify that the LED turns on and off With Snapshot active...

Страница 11: ...the image rotates to the normal orientation and automatically stops 13 Press the Thorax Surgery button repeatedly and verify that the LED turns on and off With Thorax Surgery active LED on take an x r...

Страница 12: ...n of the Generator Accuracy Verification fails to meet specifications proceed to section 5 2 Perform X Ray Calibration of the PM procedure TEST TECHNIQUE KV TOLERANCE MA TOLERANCE Manual Fluoro 60kV 5...

Страница 13: ...section are designed to ensure compliance with the 10 0 R minute entrance exposure limitation set by the Bureau of Radiological Health BRH Refer to the MA CALIBRATION section of the COMPACT 7600 Servi...

Страница 14: ...ntenance 00 878852 01 8COMPACT 7600 Pe Used and Refurbished C Arms are Available from www SharpMedical com Call us at 800 969 9800 An Independent C Arm Service Provider This PDF provided for research...

Отзывы: