Diagnostics
The following is an alphabetized list of messages that can appear on either C-Arm Control Panel display during initialization or
operation of the Series 9800 Mobile Digital Imaging System. These messages advise you of processing being done, notify you of
errors, and provide warnings to prevent personal injury or equipment damage.
4-HOUR WARM-UP REQUIRED - PRESS ANY KEY
- This message appears when the system has been inactive for more than
60 days. The four-hour delay is necessary for ion removal from the Image Intensifier. All generator functions are disabled during
this period. Press any C-Arm key to continue after the four-hour warm-up is complete.
24-HOUR RECHARGE REQUIRED - X RAYS DISABLED
- This message appears when the C-Arm batteries need recharging,
typically after long-term storage of six months or more. All generator functions are disabled during this period. Leave the
Workstation power cord plugged in, make sure the interconnect cable is properly connected, and wait 24 hours before use. It is
not necessary to turn the Workstation on.
AD CHANNEL %s FAIL – POWER OFF. WAIT 5 SECONDS
- You will see this message when an analog-to-digital channel fails
during system start-up. The actual channel number appears in place of the pound sign. Analog-to-digital channel failure causes
total shutdown of the system. First, reboot the system, leaving power off for five seconds. If the failure reappears, refer to
A/D
Channel Failure
in this section for detailed troubleshooting information.
ANODE IS HOT - XX%
- This error message appears when the X-ray tube is at 80% (+ 2%) of its rated capacity or greater. XX is
the actual percentage of capacity. The message alternates with the technique and an alarm sounds.
TO PREVENT DAMAGE TO THE X-RAY TUBE, ALLOW IT TO COOL BEFORE MAKING ANOTHER EXPOSURE. The system
blocks HLF and FILM shots. You can silence the alarm by pressing the
ALARM RESET
key.
ANODE IS WARM - XX%
- When the X-ray tube is at 70% (+ 2%) of its rated capacity, you will see this error message alternating
with the technique. XX is the actual percentage of capacity. The system displays this message as a warning only, and does not
restrict HLF or film shots. To prevent damage to the x-ray tube, allow it to cool before making another exposure.
DANGER:
Ignoring this warning can cause damage to the X-ray tube, cooling bladder, and may
result in patient burns.
BATTERY CHARGE - XX%
- The system displays this message alternating with the technique when the C-Arm batteries are at
70% to 40% of the required charge. XX is the actual charge percentage, which the system displays in the message. Fluoro, Film,
and HLF exposures are still allowed. To charge the batteries, leave the system plugged into an active AC receptacle with the
Interconnect Cable in place between the C-Arm and the Workstation.
7
Illustrated Parts
Содержание 9800C-Arm
Страница 96: ...DC Power Distribution 14 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 102: ...Battery Charging 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 114: ...Interlocks Miscellaneous None 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 128: ...Pre Charging 8 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 132: ...X ray On X ray Disable X Ray On Switch 4 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 150: ...kV Generation Location of IGBT PCB 10 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 156: ...kV Generation 16 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 170: ...mA Generation 14 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 182: ...Collimator 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 195: ...Cooling 3 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 226: ...L Arm Motion 10 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 274: ...Diagnostics All ARCNet Nodes 24 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 275: ...Diagnostics X ray Controller Node 25 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 276: ...Diagnostics Fluoro Functions Node 26 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 277: ...Diagnostics Workstation Node 27 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 278: ...Diagnostics Service Node 28 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 370: ...Calibration 86 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 395: ...Replacement Power Motor Relay PCB 25 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 453: ...Replacement Power Supply PS3 83 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...