Replacement
1. Turn the Workstation off and remove its power plug from the AC outlet.
2. Remove the C-Arm's Rear Cover and disconnect the steering linkage.
3. USING GREAT CARE TO AVOID ELECTRIC SHOCK, disconnect battery cables from P1A and P3B on Capacitor/Power
module. (Refer to Capacitor/Power module replacement procedure in this chapter for more details.)
4. Locate two battery cables just above P1A and P3B. These two cables have blue connectors that connect to each other.
USING GREAT CARE TO AVOID ELECTRIC SHOCK, disconnect battery cables from each other. (Refer to
Capacitor/Power module replacement procedure in this chapter for more details.)
5. USING GREAT CARE TO AVOID ELECTRIC SHOCK, connect a current-limiting resistor (300k or so) across
Capacitor/Power module busbars to discharge high energy capacitors inside Capacitor/Power module. (Refer to
Capacitor/Power module replacement procedure in the chapter for more details.)
3. Turn OFF CB1 on the bottom of the Capacitor/Power Module.
4. Remove the seven screws that secure the Generator Driver Cover and remove the cover.
5. Remove the Filament Driver PCB to gain access to the IGBT Snubber PCB.
6. Remove the screws that secure the wire terminals to the IGBT PCB.
7. Remove two recessed hex bolts from the IGBT PCB
8. Remove two screws from the edge of the IGBT PCB.
9. Remove the IGBT PCB and integrated heat sink.
10. If you are replacing the H.V. Tank, remove the other half of the IGBT heat sink.
70
Illustrated Parts
Содержание 9800C-Arm
Страница 96: ...DC Power Distribution 14 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 102: ...Battery Charging 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 114: ...Interlocks Miscellaneous None 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 128: ...Pre Charging 8 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 132: ...X ray On X ray Disable X Ray On Switch 4 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 150: ...kV Generation Location of IGBT PCB 10 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 156: ...kV Generation 16 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 170: ...mA Generation 14 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 182: ...Collimator 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 195: ...Cooling 3 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 226: ...L Arm Motion 10 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 274: ...Diagnostics All ARCNet Nodes 24 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 275: ...Diagnostics X ray Controller Node 25 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 276: ...Diagnostics Fluoro Functions Node 26 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 277: ...Diagnostics Workstation Node 27 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 278: ...Diagnostics Service Node 28 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 370: ...Calibration 86 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 395: ...Replacement Power Motor Relay PCB 25 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Страница 453: ...Replacement Power Supply PS3 83 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...