1-6. Check After Input Power
When Power Supply is ON, it will start setting & self-testing. Also, it maintains previous
Remote Interface Setting. Output voltage is 0V, output current is maximum.
(When the product is released, last state setting is disabled).
Read Reference "1-2. Factory mode" "Last State" Part.
Check Sequence
▌"
WWW.ODACORE.COM
" will be displayed.
If you visit our website, you will able to download manual & software, upgrade information
product information.
▌Every lamp will be lighted.
▌"
INITIALIZING
" M
ill b Di
l
d
▌"
INITIALIZING. . .
" Message will be Displayed.
While above features are displaying, a reset will be done by fixed memory.
▌Also front lamp(CV, CC, OVP, OCP, RMT, TRK, PAR) will be lighted out.
▌"
i/f - gpib
" OR "
i/f - rs232c
" will be Displayed.
It will automatically search power module when communation interface is done.
▌The Product will activate Self-Test. Descriptions are below
Front panel Test
Check whether Front panel is connected
Front panel Test
Check whether Front panel is connected
Analog Power Module Test
Check OPM Power Module
Remote Interface Test
Check Remote Interface for PC Communication
Memory Data Verification Test
Check product information & setting
ADC H/W Error Test
Check ADC HardWare Error
UnRegulated Condition Test
Check Output Voltage Floating
ADC/DAC Calibration Test
Check ADC & DAC Calibration Data
If there is an error alarm will occur and saves in fixed memory
If there is an error, alarm will occur and saves in fixed memory.
Press ERR Key to check the Error code.
Read reference "7. Error Messages" for more information.
Default Setting Value
▌OVP : OVP Maximum, OVP ON
▌Output ON/OFF : OFF
▌OCP : OCP Maximum, OCP ON
▌Remote Interface : Maintain Previous Setting
▌Output Voltage : 0V
▌
Display after Self-Testing
: "
P1-OUTPUT OFF**
"
▌Output Current : Limit Maximum Setting▌Cursor Location : Default Voltage
▌Output Select : P1(Positive channel)
Voltage : 1V Unit
▌Tracking : OFF
Current : Less than 5A, 100mA Unit
▌Parallel : OFF
Less than 50A, 1A Unit
If
did
t
d
GPIB O ti
RS232C
ill b
l
t d ft
l
d b
d
t
ill b 19200b
Note1
If user did not order GPIB Option, RS232C will be selected after released, baud-rate will be 19200bp
If user ordered GPIB Option, GPIB will be selected after released, The address is 05.
Default setting value can be different due to Last State Setting from Factory Mode.
Note2
- 11 -