22 Annex
Instruction Manual Myopia Master® (G/68100/EN Rev08)
77 / 84
Recommended separation distances between portable and mobile
RF communications equipment and the Myopia Master®,
IEC 60601-1-2:2015, table 6
The Myopia Master® in intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Myopia Master® can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Myopia Master® as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter m
Rated maximum output
power of transmitter
W
150 KHz to 80 Mhz
d= 1.2
√
P
80 MHz to 800 MHz
d= 1.2
√
P
800 MHz to 2.5 GHz
d= 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.80
3.80
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.