4 Safety Instructions
Instruction Manual Myopia Master® (G/68100/EN Rev08)
9 / 84
usefull to accomplish a test measurement of the axial length measuring
mode everyday before you start working with the Myopia Master®.
If an error occurs which you cannot correct, label the Myopia Master® as
being “out-of-order” and contact our service department, see
.
Instructions for Disassembly and Disposal
When disconnecting electrical connections, pull on the respective
plug and not on the cable itself.
Dispose of the device according to legal regulations.
Instructions on Electrical Safety
Attention
Risk of personal injury or damage to property due to an incorrect level of
safety
Connecting the Myopia Master® with its non-medical electrical
equipment (e.g. data processing equipment) to a medical electrical
system must not result in a patient safety level below that prescribed by
IEC 60601-1. If making this connection leads to the leakage current
threshold being exceeded, protective measures including a circuit breaker
must be in place.
Ensure that connections with non-medical devices are made
correctly.
Only use the power adapter listed in the packing list.
Use only a computer that meets the specifications given in this
instruction manual,
Use of a multiple socket extension cord
Risk of personal injury or material damage caused by unsafe multiple
socket extension cord
If you use a multiple socket extension cord to connect the
Myopia Master® to the power supply, you must heed the following
information:
Use an extension cord that complies with the requirements of IEC
60601-1, section 16.
Do not place the multiple socket extension cord on the floor.
Do not use more than one multiple socket extension cord.
Plug only the Myopia Master® and the computer that is being used
with the unit (if applicable) into the multiple socket extension cord.
If you are using a multiple socket extension chord it has to be supplied
with a isolation transformer.