11 Viewing the Images
32 / 68
Instruction Manual ImageCam® 3 (G/69200/XXXX/EN 0120 Rev01)
Enter the reference lengths
Open single screen mode,
.
Open the ruler image with the lowest magnification, in this case 6x
magnification.
ImageCam® 3 Program > Tools > Estimate distance
Click on the left mouse button to set the starting point in the image.
Drag the line 10 mm.
10 mm are set as the reference distance. Depending on the
magnification at which the image was captured, you can change the
reference distance to 5 mm.
Click on the white cross at one end of the line with the right mouse
button. The following window is displayed (example):
Select [new].
Fig. 11-6: Determine distance
Fig. 11-7: Selection “new”