C: Analog Output
013-20000-009-02-201510
53
The figure below shows a complete 4-20 mA loop and its 3 main components: the NeoFox unit, an
ammeter, and a voltage supply. In the diagram, the NeoFox unit is shown on the left and the customer’s
interface circuitry (voltage supply and ammeter) is on the right. Although they are drawn close together,
the NeoFox unit and the customer’s interface circuitry may be physically located far apart, connected by a
2-wire cable.
Typical 4-20 mA Interface
4-20 mA Loop Voltage Supply
As a simple current sink, NeoFox does not provide its own power for the current loop. The user must
provide an isolated power supply between 11 V and 30 V (measured at the input to the NeoFox unit) to
power the loop. This power supply
must be
completely separate from the NeoFox and its power supply;
the grounds of the Neofox and the supply
cannot be connected together
.
Most customers will use separate AC adapters to power their NeoFox unit and their interface circuitry.
Usually this will result in the two circuits floating with respect to each other so that the circuit will work
properly. However, it is important to note that NeoFox has a grounded metal case that is connected to its
circuit ground and its USB shell. Hence, customers who connect the NeoFox’s case ground to their
interface circuit’s case ground will find that the measurements are not correct. Some common situations
that could lead to unintended ground connections are the following:
Mounting both devices on a common metal surface (such as a DIN rail)
Connecting both devices to a single computer or USB hub
Using the same measurement device to measure the 0-5V output and the 4-20mA output. (The 0-
5V output is referenced to the NeoFox circuit ground.)
Using the same measurement device to communicate with NeoFox via RS232 and to measure the
output of the 4-20 mA output. (RS232 is only available in NeoFox GT models with an F or
greater in the third digit of its serial number, which are available by request).
A
Neofox
Unit
PIN 4
PIN 3
Loop
Voltage
Supply
Ammeter
Содержание NeoFox
Страница 22: ...2 Installation 14 013 20000 009 02 201510 ...
Страница 42: ...4 Using NeoFox Sport 34 013 20000 009 02 201510 NeoFox Sport Menu Menu Tree Description ...
Страница 48: ...4 Using NeoFox Sport 40 013 20000 009 02 201510 ...
Страница 50: ...A Specifications 42 013 20000 009 02 201510 ...
Страница 54: ...B Maintenance 46 013 20000 009 02 201510 ...
Страница 63: ...C Analog Output 013 20000 009 02 201510 55 ...
Страница 64: ...C Analog Output 56 013 20000 009 02 201510 ...
Страница 66: ...D Calculating Drift Per Sensor 58 013 20000 009 02 201510 ...