C: Analog Output
52
013-20000-009-02-201510
NeoFox Viewer Parameter
Potential Values
5
– Alternate O
2
Units
Current Output Source
0
– Disable analog output
1
– Oxygen
2
– Temperature (not supported)
3
– Pressure (not supported)
4
– Tau
5
– Alternate O
2
Units
Analog Voltage Lower Bound (or Voltage
Limit Minimum)
<floating point value>
- This floating point value represents
the “lower bound” of the represented range of the input
data. Must be < UBOUND and
≥ 0.
Analog Voltage Upper Bound (or Voltage
Limit Maximum)
<floating point value>
- This floating point value represents
the “upper bound” of the represented range of the input
data. Must be > LBOUND.
Analog Current Lower Bound (or Current
Limit Minimum)
<floating point value>
- This floating point value represents
the “lower bound” of the represented range of the input
data. Must be < UBOUND and
≥ 0.
Analog Current Upper Bound (or Current
Limit Maximum)
<floating point value>
- This floating point value represents
the “upper bound” of the represented range of the input
data. Must be > L
Analog Output Specifications
Name
Min
Typ
Max
Conditions
4-20 mA Loop Voltage
11 V
--
24V
R
Loop
= 2 Ω
4-20 mA Output Accuracy
--
< 160 μA
--
25°C, V
Loop
= 12V, R
Loop
= 2 Ω
0-5V Output Accuracy
--
< 50mV
--
25°C
0-5V Output Resistance
--
--
250 Ω
--
Interfacing to the 4-20mA Output
The NeoFox includes a non-isolated 4-20 mA current sink that can be used to transmit oxygen,
temperature, or tau data. It is extremely important that you set up the interface for this output correctly, or
else irreparable damage to the NeoFox unit can occur.
Содержание NeoFox
Страница 22: ...2 Installation 14 013 20000 009 02 201510 ...
Страница 42: ...4 Using NeoFox Sport 34 013 20000 009 02 201510 NeoFox Sport Menu Menu Tree Description ...
Страница 48: ...4 Using NeoFox Sport 40 013 20000 009 02 201510 ...
Страница 50: ...A Specifications 42 013 20000 009 02 201510 ...
Страница 54: ...B Maintenance 46 013 20000 009 02 201510 ...
Страница 63: ...C Analog Output 013 20000 009 02 201510 55 ...
Страница 64: ...C Analog Output 56 013 20000 009 02 201510 ...
Страница 66: ...D Calculating Drift Per Sensor 58 013 20000 009 02 201510 ...