LS-450 Series Blue LED Pulsed Light Sources 

LS-450 Series Blue LED Pulsed Light Sources 

8 

USB-LS-450 Blue LED Pulsed Light Source 

Functionally similar to the LS-450, the USB-LS-450 attaches directly to the 10-pin accessory connector of the 
USB2000 Spectrometer. Its small-footprint design combines with the portability of the USB2000 to form a 
complete, self-contained light source and spectrometer combination. 

The USB-LS-450 features the following design innovations: 

Simplified LED-Spectrometer Interface – USB-LS-450 
connects to the USB2000 Spectrometer, which connects directly 
to the USB port of any desktop or notebook PC. 

Small Footprint – USB-LS-450 and spectrometer combination 
is about the size of a deck and a half of playing cards. 

On-board Memory for O2 Measurements – FOXY Fiber Optic 
Oxygen Sensor users can conveniently store oxygen and 
temperature calibration coefficients. 

No External Power Requirements – Spectrometer and USB-
LS-450 draw power from the host PC. 

Temperature Probe Connectivity – 4-pin connector on the 
side of the USB-LS-450 allows quick interface of the USB-LS-
450-TP Temperature Probe used in the FOXY line of fiber optic 
oxygen sensors. 

USB-LS-450 Included Materials 

• 

USB-LS-450 Blue LED Pulsed Light Source 

• 

Allen wrench (to adjust collimating lens) 

USB-LS-450 Setup 

Follow the steps below to install the USB-LS-450: 

1.  Align the male 10-pin accessory input on the USB-LS-450 with the 10-pin accessory connector on the 

USB2000 Spectrometer. 

2.  Secure the USB2000 and the USB-LS-450 together by firmly interfacing the 10-pin accessory connectors. 

3.  Insert the silver thumbscrew through the interface plate of the USB-LS-450 and tighten the thumbscrew to 

secure the USB-LS-450 and the USB2000 together. 

 

USB-LS-450 (bottom) connected to the 

USB2000 Spectrometer 

Содержание LS-450 Series

Страница 1: ...d alone Version Included Materials LS 450 light source DB 15 accessory cable Power supply Connecting the LS 450 To connect the LS 450 follow the steps below 1 Plug the wall transformer into a standard...

Страница 2: ...meter See the OOIBase32 Spectrometer Operating Software Online Help System for more information on setting the integration time Note USB based spectrometers automatically flash the lamp once per milli...

Страница 3: ...p Enable option is selected in OOIBase the USB2000 spectrometer will pulse the lamp at 1 msec intervals ensuring a consistent pulse rate regardless of the configured integration time HR2000 1 Obtain a...

Страница 4: ...peration for your setup and configure your system appropriately S1 Switch The S1 switch is a three position switch on the R LS 450 The switch can be moved into the following positions Continuous wave...

Страница 5: ...88 0 976 0 2 10 Divide by 2 10 976 0 1952 0 CS Continuous Strobe Software Controlled Software Controlled Jumper Block 3 JP3 JP3 consists of two sets of pins The jumper position determines the source o...

Страница 6: ...00 You can control the pulses per second of the R LS 450 using JP3 of the S2000 ONLY if the following conditions are met The switch is turned to pulsed mode The pins in JP1 of the R LS 450 are jumpere...

Страница 7: ...times shown in the table below depending on the A D converter you are using S2000 JP3 Pin Row ADC2000 PCI Acceptable Int Times ADC1000 Acceptable Int Times 16 32 64 14 8 16 12 2 4 10 2 4 2 Integration...

Страница 8: ...ic Oxygen Sensor users can conveniently store oxygen and temperature calibration coefficients No External Power Requirements Spectrometer and USB LS 450 draw power from the host PC Temperature Probe C...

Страница 9: ...the USB LS 450 provides unaltered connections to the 10 pin accessory connector on the USB2000 Spectrometer Use this connector to configure external triggering options on the USB2000 when the USB LS...

Страница 10: ...VDC 60 mA 5 VDC Power output minimum 50 W into 600 m fiber 60 W into 600 m fiber Stability 1 0 after 2 minute warm up 1 0 after 2 minute warm up LED drive current 20 mA 5 mA switchable 20 mA 0 15 mA A...

Страница 11: ...Pull the two halves apart gently until they are approximately 1 5 apart This should expose two thin black wires that are holding the unit together These wires connect to a small socket on the LED loc...

Страница 12: ...re the anode and cathode lead pins are located The orientation of the LED should be marked with a white dot on the side of the barrel cathode side OR by differing pin lengths the short pin is the cath...

Отзывы: