LS-450 Series Blue LED Pulsed Light Sources 

LS-450 Series Blue LED Pulsed Light Sources 

2 

Using Pulsed Mode 

 

To operate the LS-450 in Pulsed mode, follow the steps below: 

1.  Connect one end of the DB-15 accessory connector to the back of the LS-450. 

2.  Connect the other end of the DB-15 accessory connector to the 15-pin connector or adapter on your 

spectrometer. For more information, see the Pulsed Mode Connection Guide section later in this 
document. 

3.  Move the switch on the back of the LS-450 to the “pulsed” setting to enable pulsed mode operation. 

4.  Configure Pulsed mode for your spectrometer.  See the Pulsed Mode Connection Guide section later in 

this document for more information. 

5.  Set the integration time in the OOIBase32 software so that a constant number of flashes occur for every 

integration cycle. If you are using the S2000 spectrometer, you will need to set the integration time to 
ensure a constant number of flashes per integration cycle. See the Pulsed Mode Connection Guide 
section later in this document for a configuration table for the S2000 spectrometer. 
 
See the OOIBase32 Spectrometer Operating Software Online Help System for more information on 
setting the integration time. 

Note:  USB-based spectrometers automatically flash the lamp once per millisecond.  Thus, there are 

always a constant number of flashes regardless of the integration time configured. 

 

Pulsed Mode Connection Guide 

Follow the instructions below for your specific spectrometer to configure Pulsed mode on the LS-450: 

S2000 

1.  Open the S2000 spectrometer housing and remove the spectrometer from the housing.  Do not tamper 

with the optical bench. 

Note:   If you have more than one channel in your system, you may have to disconnect the channels 

from one another.  The master spectrometer is always on the bottom of a multiple channel 
system. 

2.  Locate Jumper Block 3 (labeled JP3) on the green circuit board.  This jumper block consists of 10 pins, 

which are labeled by rows (/16, /14, /12, /10, and 2). 

3.  Attach the jumper to the appropriate pins to configure the number of pulses per second for your A/D 

converter. The table below illustrates the possible number of pulses per second for each A/D converter: 

S2000 JP3 Pin Row 

ADC2000-PCI Frequency (Hz) 

ADC1000 Frequency (Hz) 

/16 30.4 

15.2 

/14 120 

60.8 

/12 488 

244.0 

/10 1952 

976.0 

Controlled by flash delay in software 

 

(Continued)

Содержание LS-450 Series

Страница 1: ...d alone Version Included Materials LS 450 light source DB 15 accessory cable Power supply Connecting the LS 450 To connect the LS 450 follow the steps below 1 Plug the wall transformer into a standard...

Страница 2: ...meter See the OOIBase32 Spectrometer Operating Software Online Help System for more information on setting the integration time Note USB based spectrometers automatically flash the lamp once per milli...

Страница 3: ...p Enable option is selected in OOIBase the USB2000 spectrometer will pulse the lamp at 1 msec intervals ensuring a consistent pulse rate regardless of the configured integration time HR2000 1 Obtain a...

Страница 4: ...peration for your setup and configure your system appropriately S1 Switch The S1 switch is a three position switch on the R LS 450 The switch can be moved into the following positions Continuous wave...

Страница 5: ...88 0 976 0 2 10 Divide by 2 10 976 0 1952 0 CS Continuous Strobe Software Controlled Software Controlled Jumper Block 3 JP3 JP3 consists of two sets of pins The jumper position determines the source o...

Страница 6: ...00 You can control the pulses per second of the R LS 450 using JP3 of the S2000 ONLY if the following conditions are met The switch is turned to pulsed mode The pins in JP1 of the R LS 450 are jumpere...

Страница 7: ...times shown in the table below depending on the A D converter you are using S2000 JP3 Pin Row ADC2000 PCI Acceptable Int Times ADC1000 Acceptable Int Times 16 32 64 14 8 16 12 2 4 10 2 4 2 Integration...

Страница 8: ...ic Oxygen Sensor users can conveniently store oxygen and temperature calibration coefficients No External Power Requirements Spectrometer and USB LS 450 draw power from the host PC Temperature Probe C...

Страница 9: ...the USB LS 450 provides unaltered connections to the 10 pin accessory connector on the USB2000 Spectrometer Use this connector to configure external triggering options on the USB2000 when the USB LS...

Страница 10: ...VDC 60 mA 5 VDC Power output minimum 50 W into 600 m fiber 60 W into 600 m fiber Stability 1 0 after 2 minute warm up 1 0 after 2 minute warm up LED drive current 20 mA 5 mA switchable 20 mA 0 15 mA A...

Страница 11: ...Pull the two halves apart gently until they are approximately 1 5 apart This should expose two thin black wires that are holding the unit together These wires connect to a small socket on the LED loc...

Страница 12: ...re the anode and cathode lead pins are located The orientation of the LED should be marked with a white dot on the side of the barrel cathode side OR by differing pin lengths the short pin is the cath...