Changing a Bulb in the R-LS-450 

To change the bulb in the R-LS-450, follow the steps below: 

1.  Ensure that the unit is off and that power is disconnected from the R-LS-450. 

2.  Use the 1/20 Allen wrench to loosen the small setscrew located on the side of the blue optical housing. If 

you remove the screw fully, place it in a safe place until you replace it later in this procedure. 
 
The LED and barrel are housed within the blue optical housing. This housing also contains the SMA fiber 
connection. There are two thin black wires connected to a socket, which is in turn plugged in to the rear 
end of the LED barrel. 

3.  Slide the LED barrel out of the blue optical housing. The barrel should slide easily. If it does not, you will 

need to further loosen the setscrew (Step 2) and retry. 

4.  Remove the blue filter, seated directly in front of the blue LED. Place the filter to the side. You will only 

need this filter when using the blue LED. 

5.  Unplug the blue LED from the connection socket. Remove the socket by gently pulling the insulated black 

wire back from the rear of the barrel. The socket should unplug easily 

Note:  If you want to reinsert the default blue LED in the future, identify the original LED orientation with 

a small amount of liquid paper. 

6.  Seat the socket of the replacement LED firmly into the rear of the barrel (where the anode and cathode 

lead pins are located). The orientation of the LED should be marked with a white dot on the side of the 
barrel (cathode side) OR by differing pin lengths (the short pin is the cathode side of the LED). 
 
There is also a diagram of the LED on the circuit board of the R-LS-450, which illustrates the socket 
orientation relative to the LED. This diagram is located next to where the two black wires connect to the 
circuit board. 

7.  Return the LED barrel back into the blue optical housing of the R-LS-450. 

8.  Push the barrel all the way into the blue optical housing and tighten the setscrew on the side (loosened in 

Step 2) so that the barrel is firmly set in the housing. 

9.  Plug the unit into the 12 V power supply and turn the unit on. If the LED illuminates, installation was 

successful.  Return the R-LS-450 to the original rack unit. 
 
If the new LED does not illuminate when the switch is in the On position, consult the troubleshooting tips 
at the end of this section. 

Troubleshooting 

If the replacement LED does not illuminate when the switch is in the On position, check the following items: 

• 

Is the power supply in use correct? The light sources require a 12 V power supply with a 2.1mm center-
positive jack. 

• 

Is the power supply plugged into an outlet that is enabled and standard for the power supply you are 
using? 

• 

Is the LED in the correct orientation in respect to the socket (Anode and Cathode)? 

If these items are correct and the light source is not emitting light at the proper wavelength (or at all), please 
contact Ocean Optics Technical Support at 

[email protected]

 or at (727) 733-2447. 

Содержание LS-450 Series

Страница 1: ...d alone Version Included Materials LS 450 light source DB 15 accessory cable Power supply Connecting the LS 450 To connect the LS 450 follow the steps below 1 Plug the wall transformer into a standard...

Страница 2: ...meter See the OOIBase32 Spectrometer Operating Software Online Help System for more information on setting the integration time Note USB based spectrometers automatically flash the lamp once per milli...

Страница 3: ...p Enable option is selected in OOIBase the USB2000 spectrometer will pulse the lamp at 1 msec intervals ensuring a consistent pulse rate regardless of the configured integration time HR2000 1 Obtain a...

Страница 4: ...peration for your setup and configure your system appropriately S1 Switch The S1 switch is a three position switch on the R LS 450 The switch can be moved into the following positions Continuous wave...

Страница 5: ...88 0 976 0 2 10 Divide by 2 10 976 0 1952 0 CS Continuous Strobe Software Controlled Software Controlled Jumper Block 3 JP3 JP3 consists of two sets of pins The jumper position determines the source o...

Страница 6: ...00 You can control the pulses per second of the R LS 450 using JP3 of the S2000 ONLY if the following conditions are met The switch is turned to pulsed mode The pins in JP1 of the R LS 450 are jumpere...

Страница 7: ...times shown in the table below depending on the A D converter you are using S2000 JP3 Pin Row ADC2000 PCI Acceptable Int Times ADC1000 Acceptable Int Times 16 32 64 14 8 16 12 2 4 10 2 4 2 Integration...

Страница 8: ...ic Oxygen Sensor users can conveniently store oxygen and temperature calibration coefficients No External Power Requirements Spectrometer and USB LS 450 draw power from the host PC Temperature Probe C...

Страница 9: ...the USB LS 450 provides unaltered connections to the 10 pin accessory connector on the USB2000 Spectrometer Use this connector to configure external triggering options on the USB2000 when the USB LS...

Страница 10: ...VDC 60 mA 5 VDC Power output minimum 50 W into 600 m fiber 60 W into 600 m fiber Stability 1 0 after 2 minute warm up 1 0 after 2 minute warm up LED drive current 20 mA 5 mA switchable 20 mA 0 15 mA A...

Страница 11: ...Pull the two halves apart gently until they are approximately 1 5 apart This should expose two thin black wires that are holding the unit together These wires connect to a small socket on the LED loc...

Страница 12: ...re the anode and cathode lead pins are located The orientation of the LED should be marked with a white dot on the side of the barrel cathode side OR by differing pin lengths the short pin is the cath...

Отзывы: