LS-450 Series Blue LED Pulsed Light Sources 

LS-450 Series Blue LED Pulsed Light Sources 

1 

LS-450 Series Blue LED Pulsed Light Sources 

The LS-450 Series Blue LED Pulsed Light Sources are compact, low-cost light-emitting diodes that produce 
pulsed or continuous spectral output at 470 nm – the blue region – for high-sensitivity emission fluorescence 
measurements. 

The LS-450 light source can be combined with other sampling optics for a variety of fluorescence applications.   

The LS-450 series consists of the following light sources: 

• 

LS-450 (stand-alone version) 

• 

R-LS-450 (rack-mount version) 

• 

R-LS-450-2 (rack-mount version with 2 LEDs) 

• 

USB-LS-450 (connects directly to USB spectrometers, documented separately) 

LS-450 Stand-alone Version 

Included Materials 

• 

LS-450 light source 

• 

DB-15 accessory cable 

• 

Power supply 

Connecting the LS-450 

To connect the LS-450, follow the steps below: 

1.  Plug the wall transformer into a standard 110V outlet. 

2.  Connect the power cord to the power input on the LS-

450. 

3.  Connect an optical fiber to the SMA 905 connector on 

the LS-450. 

4.  Connect the other end of the optical fiber to the SMA 

905 connector on your spectrometer or sampling device. 

Operation 

To use the LS-450, connect the light source by following the instructions above, then proceed: 

Using Continuous Mode 

To operate the LS-450 in Continuous mode, follow the steps below: 

1.  Move the switch on the back of the LS-450 to the “contin” setting. This means that the light coming out of 

the LS-450 is continuous. 

2.  Move the switch on the back of the LS-450 to the “off” setting to turn the lamp off. 

 

LS-450 Blue LED Pulsed Light Source -

Stand-alone Version 

Содержание LS-450 Series

Страница 1: ...d alone Version Included Materials LS 450 light source DB 15 accessory cable Power supply Connecting the LS 450 To connect the LS 450 follow the steps below 1 Plug the wall transformer into a standard...

Страница 2: ...meter See the OOIBase32 Spectrometer Operating Software Online Help System for more information on setting the integration time Note USB based spectrometers automatically flash the lamp once per milli...

Страница 3: ...p Enable option is selected in OOIBase the USB2000 spectrometer will pulse the lamp at 1 msec intervals ensuring a consistent pulse rate regardless of the configured integration time HR2000 1 Obtain a...

Страница 4: ...peration for your setup and configure your system appropriately S1 Switch The S1 switch is a three position switch on the R LS 450 The switch can be moved into the following positions Continuous wave...

Страница 5: ...88 0 976 0 2 10 Divide by 2 10 976 0 1952 0 CS Continuous Strobe Software Controlled Software Controlled Jumper Block 3 JP3 JP3 consists of two sets of pins The jumper position determines the source o...

Страница 6: ...00 You can control the pulses per second of the R LS 450 using JP3 of the S2000 ONLY if the following conditions are met The switch is turned to pulsed mode The pins in JP1 of the R LS 450 are jumpere...

Страница 7: ...times shown in the table below depending on the A D converter you are using S2000 JP3 Pin Row ADC2000 PCI Acceptable Int Times ADC1000 Acceptable Int Times 16 32 64 14 8 16 12 2 4 10 2 4 2 Integration...

Страница 8: ...ic Oxygen Sensor users can conveniently store oxygen and temperature calibration coefficients No External Power Requirements Spectrometer and USB LS 450 draw power from the host PC Temperature Probe C...

Страница 9: ...the USB LS 450 provides unaltered connections to the 10 pin accessory connector on the USB2000 Spectrometer Use this connector to configure external triggering options on the USB2000 when the USB LS...

Страница 10: ...VDC 60 mA 5 VDC Power output minimum 50 W into 600 m fiber 60 W into 600 m fiber Stability 1 0 after 2 minute warm up 1 0 after 2 minute warm up LED drive current 20 mA 5 mA switchable 20 mA 0 15 mA A...

Страница 11: ...Pull the two halves apart gently until they are approximately 1 5 apart This should expose two thin black wires that are holding the unit together These wires connect to a small socket on the LED loc...

Страница 12: ...re the anode and cathode lead pins are located The orientation of the LED should be marked with a white dot on the side of the barrel cathode side OR by differing pin lengths the short pin is the cath...

Отзывы: