MNL-1002 Rev A
14
NIRQuest512
NIRQuest512-1.9 NIRQuest512-2.2
NIRQuest512-2.5
NIRQuest256-2.1
(environmental):
TEC range:
20 -- 35ºC below
ambient
30 -- 50ºC below ambient
TEC stability:
+/-0.5 ºC of set temperature in <1 minute; typical long-term sta/-0.1 ºC
COMPUTER
Computer
interfaces:
USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud (custom configuration)
Peripheral
interfaces:
I2C inter-integrated circuit
1
“Useable range” is defined in the context of the NIRQuest
model’s detector response and its typical grating response. For example, the 512
-element detector has response at 850 nm,
but grating response begins at 900 nm. The G9206 256-element detector response is sensitive to TEC temperature, and has response only to 2050 nm when the TEC is set to -20 ºC.
The G9208 256-element and 512-element detector has response to 2550 nm, but the grating efficiency drops off at 2500 nm.
2
Other filter options are available for order-sorting in the NIRQuest 512-2.5. NIRQuest 512-2.5 ships with a 2
nd
-order filter. See an Applications Scientist for details.
3
Optical resolution (FWHM) depends on grating and slit selection.
4
SNR will decrease at longer integration times.
5
Maximum integration times are defined as the longest amount of time one can integrate the spectrometer before the dark level rises to half of full scale.