NEP-9500 MANUAL V1.01
Observator Instruments V1.01
Copyright 2017
©
Page 10
5.
PINOUT & WIRING DIAGRAM
Because long cable lengths are common some electrical factors should be taken into account and
these are discussed below.
There are two wiping protocols available on the NEP-9500 probes:
1. A single wipe externally initiated by momentarily reseting the probe (specify
during the ordering).
2. A periodic automatic wipe (specify during the ordering).
Voltage Output version
ANALITE NEP-9500 probes may have an analogue voltage output with a nominal
output impedance of 500 ohms. For particularly long cable runs, the 4-20mA current
output version is recommended.
Wire color
Voltage
option
Green
Ground (0V Common)
Shield
Shield (No connection within probe)
Brown
8 to 30V DC power supply (+Ve)
Grey
Wiper (Probe sleep/Wipe)
Yellow
NTU
Voltage
Common
White
NTU
Voltage
Output