background image

               

                          

NEP-9500 MANUAL V1.01

                     

 

   

 

 

Observator Instruments V1.01  

     Copyright 2017 

©

 

 

Page 15 

8.

 

INSTALLATION 

 

They are aspects to consider when preparing to install the ANALITE NEP-9500 series probes on the 
field: 
 

1.  Installation of the probe proper into the environment where measurements are to take place. 
2.  Physical cabling considerations. 
3.  Electrical cabling considerations. 

 

Install the probe for the first time 

 

The probe is normally installed with the optics pointing downwards (but not vertical) or in a horizontal 
alignment. In a simple application the probe is simply immersed into the water to the desired depth, 
but  within  the  depth  rating  of  the  probe.  Please  note  the  depth  rating  is  based  on  static  water. 
Allowances  must  be  made  for  the  effect  of  flowing  water  to  ensure  the  static  depth  rating  is  not 
exceeded. If the probe is to be installed downwards then it is recommended to install it a few degrees 
away from the vertical to allow bubbles to escape away from the optic face.  
 
Probes with integrated wiping should be installed such that they can be easily retrieved from time to 
time to replace the wiper arm assembly. It is important that the optic end of the probe is kept clear of 
obstruction  such  as  the  river  bed. The  minimum  distance  between  the  optic  head  and  any  object 
should be 50mm (2”). 
 
The standard casing of the ANALITE NEP-9500 series probes are made of 6 stainless steel but the 
optic  face  is  made  of  plastic  materials  and  so  should  be  protected  from  accidental  scratching  or 
abrasion. The wiper arm assembly should be replaced periodically to avoid abrasive material build-
up in the pad that may eventually abrade the optic face. The optic face is partially protected from 
damage by the protruding castellation in the probe casing. 
 
To avoid crevice corrosion on the 6 stainless steel casing in salt and acidic water deployments it is 
strongly recommended that the probe be thoroughly washed in clean water after deployment and 
prior to storage. Failure due to crevice corrosion is not covered under warranty. A composite cased 
version  is  available  for  corrosive  and  sea  water  environments  but  its  pressure  rating  is  limited  to 
30m. 
 
If the probe is to be installed in a glanded fitting (for insertion into a pipe etc.) then care must be 
taken to ensure the sealing surface pressures offered by the gland fitting are not excessive so as to 
not cause distortion of the probe casing and force leakage. The ANALITE NEP-9500 series probes 
are  thin  wall  instrument  and  so  glanding  pressure  must  be  minimal  and  spread  over  the  largest 
possible area. Do not cut or damage the outer sheath of the cable. Water may enter the probe through 
holes or cuts in the cable sheath. Where damage may occur due to river rocks striking or rolling over 
the  probe  body,  a  protective  shroud  should  be  used  which  can  be  made  of  simple  PVC  piping  or 
stainless  steel  whichever  the  situation  warrants.  A  shroud  is  available  for  the  NEP-9500  as  an 
accessory under the part number NEP-9SHRD. Such a shroud not only protects the probe but also 
assist in maintaining a minimum distance between probe optics and any local obstructions.  

Содержание NEP-9500 Series

Страница 1: ... PROBE User Manual Version 1 01 Oct 2017 Author Observator Instruments NEP 9500 SERIES TURBIDITY PROBE ABN 56 007 283 963 8 10 Keith Campbell Court PO Box 9039 Scoresby VIC 3179 AUSTRALIA Tel 61 3 8706 5000 Fax 61 3 8706 5049 Email info au observator com Web www observator com ...

Страница 2: ...nts for more information Document Name Download from Author Online Training documents http download observator com files dir User manuals NEP 9500 Ludovic Grosjean Latest NEP5000 Manual http download observator com files dir Software NEP50xx Provides an in depth information of the configuration and calibration Niran Pelpola Distribution list This document has been distributed to Name Company Posit...

Страница 3: ...ing considerations 12 6 DESCRIPTION 13 7 ACCESSORIES 14 8 INSTALLATION 15 Install the probe for the first time 15 Physical cabling considerations 16 Miscellaneous 16 9 DEPLOYMENT 17 Installation 17 Deploy the probe 18 Retrieve the probe 19 10 MAINTENANCE 20 Wiper replacement 20 Cleaning the sensor 22 11 ELECTRICAL CONFORMITY 23 EC Declaration of Conformity according to Council Directive 89 336 EEC...

Страница 4: ...NEP 9500 MANUAL V1 01 Observator Instruments V1 01 Copyright 2017 Page 4 Page intentionally left blank ...

Страница 5: ...400NTU 1 000NTU 3 000NTU and 5 000NTU Specifically the ANALITE NEP 9500 probes are designed for applications that will not allow bio fouling to build up such as short monitoring deployment or placement in fast and cold running water The ANALITE NEP 9500 probes however with their integral wiper assembly are designed where bio fouling or sedimentation buildup is likely The ANALITE NEP 9500 series of...

Страница 6: ...rial process monitoring 8 Sludge and dredge monitoring 9 Sediment load monitoring Which model is best used is dependent on the application the measuring environment the logging equipment and the monitoring period deployment times required The ANALITE turbidity probes are not suitable in situations where they may be abraded by large particles such as sand and under these circumstances the reading m...

Страница 7: ... functioning of Observator NEP 9500 the sensors must be installed according installation instructions Never obstruct the wiper or prevent the wiper from wiping Always clean and dry the sensor after use After end of life dispose this product according local regulations or return to manufacturer ...

Страница 8: ...servator Instruments V1 01 Copyright 2017 Page 8 4 SPECIFICATION Glanded cable option default Turbidity sensor Dimensions Length 320 4mm Diameter 31 70mm Subconn option Turbidity sensor Dimensions Length 297mm Diameter 31 7mm ...

Страница 9: ...0 C to 50 C Wiper Wiper Arrangement NEP 9500 SERIES only Disposable and consumable Foam Pad on PVC or Acetal arm Field replaceable Mounted on central shaft fixed by hex set screw onto flat of wiper shaft Actuation NEP 9500 SERIES only 1 By external TTL CMOS active low pulse or momentarily contacting the wiper actuation conductor pin 5 to 0V pin 3 2 Automatic wipe periodic and power on Please speci...

Страница 10: ...omentarily reseting the probe specify during the ordering 2 A periodic automatic wipe specify during the ordering Voltage Output version ANALITE NEP 9500 probes may have an analogue voltage output with a nominal output impedance of 500 ohms For particularly long cable runs the 4 20mA current output version is recommended Wire color Voltage option Green Ground 0V Common Shield Shield No connection ...

Страница 11: ...below For very long cable runs 200m it may be advisable to connect the NTU Return conductor to the Supply Common conductor at the end of the cable to reduce the voltage drop along the run When a NEP 9500 series probe is ordered with the 4 20mA current output the probe is configured as a high current source The load must be less than 270 ohms to ensure proper operation over the probe s specified po...

Страница 12: ...be considered given that many installations are in the wild It is important to note that the NEP 9500 series probes have the stainless steel casing terminated only to the cable shield There is no electrical contact between the casing terminated only to the cable shield There is no electrical contact between the casing and the probe electronics The signal ground 4 and power ground 3 are electricall...

Страница 13: ... you will find in the box Product identification Each product code can be used to identify each product specification and determine which wiring must be used for the configuration as detailed in the examples below Items found in the box Figure 1 1 NEP 9500 Probe NEP CBL XX Cable Length XX in meters to be specified at time of order ...

Страница 14: ...17 Page 14 7 ACCESSORIES Observator Instruments offers a wide range of accessories for NEP 9500 range of product directly available from the website Accessories Figure 2 1 Case Figure 2 2 Wiper replacement kit Figure 2 3 Shroud NEP 9SHRD Stainless protective shroud ...

Страница 15: ...d from accidental scratching or abrasion The wiper arm assembly should be replaced periodically to avoid abrasive material build up in the pad that may eventually abrade the optic face The optic face is partially protected from damage by the protruding castellation in the probe casing To avoid crevice corrosion on the 6 stainless steel casing in salt and acidic water deployments it is strongly rec...

Страница 16: ... must be taken to ensure the connectors are properly mated to ensure a seal yet not screwed together too tightly so as to make their disconnection difficult after a long deployment Prior to mating make sure both the probe and cable connectors are dry in the termination area otherwise erratic operation may occur due to moisture Miscellaneous The wiper on the NEP 9500 probes can be actuated by momen...

Страница 17: ...ements Typical example involves the use of an extension cable for Subconn connector Option which can be purchased separately Connect the Subconn Female extension cable to the probe according to the matching pin configuration as shown below Make sure the connector is sealed by screwing tight the connector Install the probe to the mounting location ready for deployment ...

Страница 18: ...ring appropriate safety equipment in safe operation conditions 2 The probe is normally installed with the optics pointing downwards but not vertical or in a horizontal alignment 3 In a simple application the probe is simply immersed into the water to the desired depth but within the depth rating of the probe Please note the depth rating is based on static water ...

Страница 19: ...ptic face is partially protected from damage by the protruding castellation in the probe casing If the probe body is to be installed in a glanded fitting for insertion into a pipe etc the care must be taken to ensure the sealing surface pressures offered by the gland fitting are not excessive so as to not cause distortion of the probe casing and force leakage The NEP 9500 probes are thin wall inst...

Страница 20: ...the wiper pad to determine if the material is deteriorating or is impregnated with material from bio fouling Normally the wiper should be replaced every 4 to 6 weeks of deployment In addition as a precaution we recommend changing the wiper prior each deployment The wiper is a consumable item and a spare is provided with each NEP 9500 series probe along with a hex key to loosen and fasten the wiper...

Страница 21: ...y one half its original thickness Tighten the set screw do not over tighten It is important that the wiper arm does not make contact with the probe face only the pad should be in contact A gap of 0 5mm between the wiper arm and the optic face is typical when a new pad has been properly installed Note It is imperative that the set screw be fastened squarely aligned onto the flat on the shaft otherw...

Страница 22: ...recommended that the probe be thoroughly washed in clean water after deployment and prior storage On the field wash the probe with fresh water and clean it with a soft cloth In the office we recommend to clean the sensor with isopropyl alcohol available from grocery store and dry the sensor with compressed air ...

Страница 23: ...ruments Pty Ltd To which this declaration relates are in conformity with the protection requirements of Council Directives 89 336 EEC on the approximation of the laws relating to electromagnetic compatibility This Declaration of Conformity is based upon compliance of the product with the following harmonized standards Emissions EN50081 1 1992 Immunity EN50082 1 1997 Signed by Dana Galbraith Managi...

Страница 24: ...ndition for details Telephone Technical Support You can contact us any time about Technical Support Repair and Service Coverage Please enquire about a maintenance program If warranty or repair is required please request an RMA number at the Observator Website https observator com en support rma request You will need 1 Your Serial Number can be found on the probe sticker 2 A valid E Mail S N 108 22...

Страница 25: ...stributor from whom you purchased the probe or our Service Centre If the probe is being returned for service under warranty please supply proof of purchase and the Warranty document which has been emailed to you during the warranty registration All support enquiries must include 1 The serial number of the probe labelled on the casing of the probe 2 The shipping address for probe returns Contact mo...

Отзывы: