The O

akworks

®

 Spine Positioning System is a patient cradle device for use in 

diagnostic and therapeutic procedures of the spine. It is intended to be used solely for 

the purpose of aiding in patient positioning and comfort during non-surgical 

imaging or spinal injection procedures. It may  also be used during minimally 

invasive surgical procedures such as vertebroplasty or kyphoplasty. The Spine 

Positioning System, its secondary components and optional components are 

suitable for use in fluoroscopy suites.

t

 introduction

The 

Spine Positioning System

 is an integral component of  the pain management 

fluoroscopy suite. With this system procedural set up time is reduced, patient comfort 

is enhanced and unwanted movement is minimized.  Most importantly, the target anat-

omy is more readily visualized  which allows the physician to perform spine procedures 

in a more efficient and secure manner. In collaboration with leading pain management 

physicians, Oakworks designed the Spine Positioning System in an effort to achieve the 

critical balance between optimal imaging and patient comfort. The radiolucent adjustable 

frame and versatile padding system provide a metal free imaging support platform capable 

of quickly positioning a wide variety of patient physiques for extended periods of time. 

The adjustable face rest position provides individualized positioning for all types of cervical 

procedures and anatomy. The contoured torso support pad is complimented by a host of  

uniquely shaped and sized adjuster pads and wedges that enable a multitude of positioning 

combinations for ideal patient comfort and imaging needs for all spinal column procedures.  

Warning

Warnings

A patient safety strap is required during all procedures. Follow normal and required 

safety protocol for all procedures where the patient is in an elevated position for the 

procedure (straps, attendants, etc.). Always be certain that attending staff is aware of 

the patient’s position while the device is in use. Reposition the patient if necessary 

to promote stability. Due to the increased distance between the patient and the table 

surface, additional safety measues are recommended when the table top is not used 

in a level position due to the risk of the patient falling off the table.

The O

akworks

®

 

Spine Positioning System

 is not designed for use with diagnos-

tic x-ray systems where the x-ray generator is located above the radiographic table 

and the film cassette or image intensifier is located below the radiographic table. 

The X-Ray generator 

muSt be

 located below the radiographic table. The Spine 

Positioning System is not designed for use with magnetic resonance imaging sys-

tems. The Spine Positioning System is not intended for use in cranial procedures.

Do not overhang the platform frame beyond the warning line on the frame.

Operate the C-arm of the fluoroscopy system with the Spine Positioning System 

in place before using the device with a patient for the first time. Make sure there is 

adequate clearance to permit free C-arm rotation for both the patient and the posi-

tioning device.

Do not permit the patient to push down on the Crescent Face Cushion in an effort 

to lift themselves up while dismounting the platform and/or the  

fluoroscopy table.

The Spine Positioning System should generally not be used when a patient is under general 

anesthesia, especially when prolonged cases are performed. This will reduce  

the risk of ocular or facial nerve injury.

Содержание Spine Positioning System

Страница 1: ...so Support Pad 2 7 x 12 rectangular adjuster pads 2 8 x 16 rectangular adjuster pads 2 Contoured Torso Wedges pictured here with small wedge 1 Radiolucent frame with adjustable face rest 1 carry case...

Страница 2: ...t to change without notice and should not be construed as a commitment by Oakworks Inc Oakworks Inc assumes no responsibility for any errors that may appear in this document nor does it make expressed...

Страница 3: ...ns 2 3 Symbol Identification 4 Description of Components 5 7 Directions for Use 8 15 Maintenance Service 17 21 cleaning disinfecting 17 maintenance 18 19 list of parts 20 specifications 21 warranty 21...

Страница 4: ...omfort and imaging needs for all spinal column procedures WARNING warnings A patient safety strap is required during all procedures Follow normal and required safety protocol for all procedures where...

Страница 5: ...l cam locks are secure before relinquishing support of the positioning assembly The Cervical Support System see page 5 has metal parts that can cause back scatter of x rays When x rays are present wea...

Страница 6: ...mbol when used in this manual or on product labels indicates that the product should be protected from moisture The humidity specifications for Transport Storage are listed on page 23 This symbol when...

Страница 7: ...Crescent face rest pad One cam lock facilitates cervical flexion and extension 1 Crescent Face Rest Pad PN EM2410 SP 1 Contoured Torso Support Pad pn 0546 06 2 7 x 12 rectangular adjuster pads pn 4402...

Страница 8: ...enhance patient stability by reducing sway while in the device 0550 06 small CONTOURED Torso WEDGE 0554 06 large CONTOURED Torso WEDGE The Contoured Torso Wedge is constructed of dense foam and cover...

Страница 9: ...gular Adjuster pads are used either individually or combined to reduce lumbar lordosis and or increase chest height to allow for shoulders to natural ly decend out of the plane of the cervical and tho...

Страница 10: ...trap is recommended during all procedures A patient saftey strap must be used during procedures where lateral roll function is used Directions for use 8 Directions for Use face rest platform adjustmen...

Страница 11: ...e face rest and rotate the facerest flat against the base frame This will protect the facerest support platform during transport When placing the Spine Positioning System in the bag put some pads wedg...

Страница 12: ...ch cervical procedures Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small Contoured Torso Wedge 1 7 x 12 Rectangular Adjuster Pad 1 8 Semi Round Bolster 1 Dir...

Страница 13: ...figured with two extra adjuster pads under the lumbar spine as a method to reduce lumbar lordosis and enhance comfort AP image of the lumbar spine Components used Platform Frame 1 Crescent Face Cushio...

Страница 14: ...e T1 2 interlaminar space Contralateral oblique showing the upper thoracic facet joints Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small Contoured Torso Wed...

Страница 15: ...age through the C1 2 segment Lateral image through the C1 3 segments Mary AP image through the C1 2 joints Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small...

Страница 16: ...n Right thoracic oblique image to visualize the trajectory for a transforaminal injection Contralateral oblique showing the trajectory for targeting the mid thoracic facet joint Components used Platfo...

Страница 17: ...ncluding the C7 T1 segment Directions for Use Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small Contoured Torso Wedge 1 7 x 12 Rectangular Adjuster Pad 1 8 x...

Страница 18: ...edures such as interlaminar epidural steroid injections facet injections medial branch blocks and medial branch radiofrequency neurotomy Components used Platform Frame 1 Crescent Face Cushion 1 Contou...

Страница 19: ...Virox Disinfecting All components may be disinfected using a 10 sodium hydrochlorite bleach solution Be sure to to remove excess solution and dry thoroughly All fabric covered components may be disinf...

Страница 20: ...d Replace any damaged or worn components Inspect face rest platform locking mechanism weekly Use the following procedure Step 1 Lock the platform cam Step 2 Rock platform up down Gently rock platform...

Страница 21: ...hanism is to simply put a thin piece of paper between the 2 pivot parts avoiding the pins and pull on the paper If the paper pulls out cam tightening is required Maintenance Cam tightening if required...

Страница 22: ...ame 1 EM2410 SP Crescent Face Cushion 1 0546 06 Contoured Torso Support Pad 1 0550 06 Small Contoured Torso Wedge 1 0554 06 Large Contoured Torso Wedge 1 4402 06 7 x 12 Rectangular Adjuster Pad 2 0552...

Страница 23: ...Pad 1 5 5 x 16 x 27 0550 06 Small Contoured Torso Wedge 1 2 5 x 16 x 23 0554 06 Large Contoured Torso Wedge 1 3 5 x 16 x 29 4402 06 7 x 12 Rectangular Adjuster Pad 2 1 x 7 x 12 0552 06 8 x 16 Rectangu...

Страница 24: ...rch 2010 Edition 2 May 2010 English Printed in U S A Contact Information Oakworks Medical 923 East Wellspring Road New Freedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com European...