CLEANING & DISINFECTION

RECOMMENDED CLEANERS/DISINFECTANTS

To reduce the risk of electric shock:

-  Always unplug this furnishing from the electrical outlet before cleaning.

DANGER

Reference the Recommended Cleaners and Disinfectant list (MMINML0008-EN) that came with the 

table. This information can also be found at www.oakworksmed.com under product information.  
All cleaners and disinfectants have the ability to degrade the upholstery to some extent. However, fol-

lowing the recommended cleaner and disinfectant list and cleaning process will provided the best care 

for your table and support a long product life. 
OAKWORKS® recommends a prepackaged wipe for cleaners/disinfectants to ensure best distribution 

of disinfectant for the required kill time, without leaving excess residue and/or overexposing compo-

nents therefore minimizing the potential for damage to materials. Please read and follow disinfectants 

manufacturers’ directions for cleaning and disinfection.
OAKWORKS® does NOT recommend the use of cleaners/disinfectants containing Hydrogen Peroxide, 

Acetic Acid, or Phenolics. These chemicals can cause damage to the appearance and/or material 

integrity of various components. Also, while the recommended cleaners/disinfectants list includes 

products containing Quaternary Ammonium compounds (“quats”), not all products containing quats 

are approved for use. Some contain additional detergents and/or surfactants which can damage some 

materials. 
A note on Bleach: While a 10% sodium hypochlorite (household bleach) solution (EPA No.: 5813-100 

or equivalent) can be an effective disinfectant and is dilute enough to be benign to most materials, it 

alone is not an effective cleaner and a separate product must be used for the initial cleaning steps of 

the procedure. Because of possible chemical incompatibilities between various cleaning products and 

bleach, utmost care must be taken by the user to avoid potential exposure to harmful or toxic by-prod-

ucts of the combination. Also, because bleach leaves a potentially corrosive residue as it evaporates, 

it must be rinsed with clean water after disinfection.

Use of non-approved cleaners or disinfectants may lead to damage to upholstery and other 

materials found on the table and will void the warranty. 

CLEANING PROCESS

Follow the cleaners/disinfectant manufacturers’ directions for use. Please note that cleaning and disin-

fecting an OAKWORKS® table is a two part process. First it must be cleaned of any visible soil, then it 

can be disinfected. OAKWORKS® recommends that the table be positioned in the flat position during 

the cleaning process. Please follow this procedure for best results:

1.  Using an approved cleaner or mild liquid soap and water, clean any visible soil off of the table, 

working from the top to the bottom of the table. It is recommended that the upholstery be 

cleaned at least once a week to prevent disinfectant build up. 

2.  Rinse with clean water and dry with a clean cloth or towel. 
3.  Using an approved disinfectant, thoroughly disinfectant all surfaces of the tabletop and any high-

contact areas such as handles, handsets, etc., making sure they remain wet for the disinfectant 

manufacturer’s recommended contact time. Do not allow disinfectant to pool on the upholstery 

after the recommended contact time. 

4.  Wipe off any excess liquid with a cloth or towel and clean water. 
5.  Dry all surfaces with a clean cloth or towel. 

Avoid using writing instruments or other similar instruments around the upholstery as it can cause per-

manent staining. If this does occur, do not wipe with an alcohol based cleaner. Instead, blot the stain 

with a clean cloth/ paper towel. Use a 10% bleach dilution to remove the stain. Follow this with a rinse 

of clean water. 

Содержание CFPMFXH

Страница 1: ...p CFPMFXH Integrated headrest top Rectangular top CFPM100 Integrated headrest top Rectangular top CFPM300 Rectangular top CFPM301 Integrated headrest top CFPM301 Rectangular top CFPMB301 Rectangular t...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Страница 3: ...le End Extensions 22 Lithotripsy Table used as Urology Table 22 Carbon Fiber Arm Board 23 Adjustable T Rail Clamp 23 Spine Positioning System II 24 Fluoro Extender 24 Reference Usable Imaging Area Dia...

Страница 4: ...orienting the equipment SYMBOL IDENTIFICATION This symbol when used in this manual and on product labels represents a caution warning Be sure to read and comply with all precautions and warnings This...

Страница 5: ...is in motion Be certain that the table is completely lowered prior to discharging an ambulatory patient The patient may lose balance and fall NOTE The CFPMFXH is a fixed height table It is recommended...

Страница 6: ...of 1 10 mm as measured at 100 kVp and a half value HVL of 3 6 mm NOTE Bariatric Top has a typical aluminum filtration equivalence of 1 50 mm as measured at 100 kVp and a half value HVL of 3 6 mm Oakw...

Страница 7: ...ing Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 368 lbs 167 kg OPTIONS 5 13cm Locking Casters 1 2 5cm...

Страница 8: ...sters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 343 lbs 156 kg O...

Страница 9: ...m Locking Casters Locking Unlocking Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all pr...

Страница 10: ...g Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 498 lbs 226...

Страница 11: ...10 cm Locking Casters Locking Unlocking Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during a...

Страница 12: ...40 kg distributed evenly 4 10 cm Locking Casters Locking Unlocking Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint S...

Страница 13: ...ile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 548 lbs 249 kg OPTIONS 5 13cm...

Страница 14: ...iolucent dimensions 19 48cm W x 23 7 60cm Table Extensions 1 23 58cm W x 28 71cm L End 1 23 58cm W x 16 41cm L OR 23 58cm W x 20 51cm L Short Table Voltage Options 120V 60Hz or 230V 50Hz 2 8 20cm W x...

Страница 15: ...cm High US Standard Urology Bag Hoop Clips onto T rails bag not included Table Capacity 500 lbs 227 kg distributed evenly 200 lbs 91 kg on all extensions 4 10 cm Locking Casters Locking Unlocking Cast...

Страница 16: ...Clips onto T rails bag not included Table Capacity 500 lbs 227 kg distributed evenly 200 lbs 91 kg on all extensions 4 10 cm Locking Casters Locking Unlocking Casters for mobile or permanent position...

Страница 17: ...oduct is equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropriate outlet that is properly installed and grounded in accordance with...

Страница 18: ...ON CAUTION 15 CAUTION Do not apply excessive force to the foot control as it may damage the control Do not use the foot control if signs of wear or damage are visible as this can cause unintentional m...

Страница 19: ...ateral Tilt OR Vertical Height Trendelenburg Longitudinal Motion Lateral Motion Control functions dependent upon table s fea tures Test each of the table s functions prior to use with a patient CAUTIO...

Страница 20: ...e patient from falling off the table The strap has hook and loop fasteners attached to the ends The strap is attached as follows 1 Place the center of each strap around the patient at two places Reloc...

Страница 21: ...nchored Replace pad by first centering on the table and then pressing flaps in place anchoring with the hook and loop fasteners ADJUSTING THE CRESCENT FACE PADS INTEGRATED HEADREST TOPS ONLY PRONE SUP...

Страница 22: ...the red button To reset rotate the red button clockwise If necessary you can also remove the plug from the outlet Be sure access is not blocked for disconnecting the table from power 19 LOCKING THE CA...

Страница 23: ...e warning line on the tabletop Casters should be locked when patient is mounting or dismounting the table Table should be flat when mounting and dismount ing Patient should mount and dismount near the...

Страница 24: ...Extensions are clamped to the T rail with a locking clamp design There are 2 lateral extensions for the Urology Table 2 5 x 24 13 x 61cm These extensions can be used as required to extend the effectiv...

Страница 25: ...28 58 x 71cm and 1 23 x 16 58 x 41cm Lithotripsy Urology 1 23 x 28 58 x 71cm Urology LITHOTRIPSY TABLE USED AS A UROLOGY TABLE If the extensions are removed from the Lithotripsy Table the table can b...

Страница 26: ...ion of the Arm Board Fold the Base and Support sections for storage Prone Position at 135 Supine Position at 180 Support Section Base Section T RAIL CLAMP USE ACCESSORY 1 The clamp must be installed o...

Страница 27: ...m LARGE RECTANGULAR ADJUSTER PAD 7 X 12 X 1 18 x 31 x 4cm SMALL RECTANGULAR ADJUSTER PAD 8 20cm SEMI ROUND BOLSTER FLUORO EXTENDER ACCESSORY CAUTION Do not place undue weight or downward pressure on t...

Страница 28: ...REFERENCE USABLE IMAGING AREA CFPMFXH CFPM100 INTEGRATED HEADREST TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE 25...

Страница 29: ...26 REFERENCE USABLE IMAGING AREA CFPMFXH CFPM100 RECTANGULAR TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE...

Страница 30: ...REFERENCE USABLE IMAGING AREA CFPM400 CFPM300 CFPM401 CFPM301 INTEGRATED HEADREST TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE 27...

Страница 31: ...28 REFERENCE USABLE IMAGING AREA CFPM400 CFPM300 CFPM401 CFPM301 RECTANGULAR TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE...

Страница 32: ...LIQUE 30 VIEW FROM VERTICAL CAUDAL 30 VIEW FROM VERTICAL 47 0 119 38cm No Image 23 3 59 06cm 15 5 39 37cm 21 0 53 34cm 2 8 6 99cm TYP 47 0 119 38cm No Image 20 0 50 80cm 19 0 48 26cm 23 0 58 42cm 2 0...

Страница 33: ...IQUE 30 VIEW FROM VERTICAL CAUDAL 30 VIEW FROM VERTICAL 47 0 119 38cm No Image 23 3 59 06cm 15 5 39 37cm 21 0 53 34cm 2 8 6 99cm TYP 47 0 119 38cm No Image 20 0 50 80cm 19 0 48 26cm 23 0 58 42cm 2 0 5...

Страница 34: ...ocedure Because of possible chemical incompatibilities between various cleaning products and bleach utmost care must be taken by the user to avoid potential exposure to harmful or toxic by prod ucts o...

Страница 35: ...nds control and foot control cables and all visible wiring Check the hand foot control for signs of wear or damage replace if necessary Visually inspect components for obvious damage that could cause...

Страница 36: ...window of the Product Label CFPM400 Table Manufacturer Oakworks Inc 923 East Wellspring Road New Freedom PA USA 17349 1 717 235 6807 www oakworks com GTIN SN 00817463021325 QTY 1 EA Manuf date 2016 1...

Страница 37: ...andards 21 CFR Sub chapter J in effect at the time of manufacture for radiographic tables ENVIRONMENTAL CONDITIONS Conditions Temperature Humidity Atmospheric Pressure Normal Use 50 10 C to 104 40 C 2...

Страница 38: ...upporting The table is intended for use in the electromagnetic environment in which radiated RF disturbances are con trolled The customer or the user of the table can help prevent electromagnetic inte...

Страница 39: ...he applicable regulations and requirements of the local state environmental protection agency state health agency Centers for Disease Control CDC Occupational Safety and Health Administration OSHA U S...

Страница 40: ...USER MANUAL FCC Rules and Regulations Title 47 Part 15 Subpart B Class A Unintentional Radiators NOTE This equipment has been tested and found to comply with the limits for a Class A digital device pu...

Отзывы: