2

IMPORTANT SAFETY INSTRUCTIONS

CAUTION

READ AND SAVE THESE INSTRUCTIONS

IMPORTANT SAFETY INSTRUCTIONS

After unpacking your table, inspect it thoroughly for damage. If you suspect a problem do not use the table 

and call our customer service department at 717.235.6807.
Improper use of this device can cause injury. Be sure to read all operating instructions prior to use.
Weight limit (patient and accessories): 227 kg / 500 lbs. distributed evenly. (Bariatric Table only: 340 kg / 

750 lbs.)
Do not sit beyond warning line on carbon fiber top. On Lithotripsy / Urology Top - Do not sit on any exten-

sions (Weight Limit: 200 lbs. / 91 kg.)
The table utilizes four locking casters to permit movement of the table within the imaging suite.  Accidental 

movement of the table may occur. Lock casters prior to accomplishing imaging of the patient. Lock casters 

when the patient is mounting & dismounting the table.
The table should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is 

necessary, the table should be observed to verify normal operation in the configuration in which it will be used. 
The use of Accessories, transducers, and cables other than those specified (with the exception of transducers 

and cables sold by the Manufacturer of this device as replacement parts for internal components) may result in 

increased Emissions or decreased Immunity of the table.
The table is designed to be a standalone table used with radiographic equipment. This table must not be 

modified or incorporated into any other equipment.
There may be pinch points under the table top. Do not have any body part under the top when the table is 

in motion.
Be certain that the table is completely lowered prior to discharging an ambulatory patient. The patient may lose 

balance and fall. NOTE: The CFPMFXH is a fixed height table. It is recommended that all patients are assisted 

onto the table (physical assistance & step stools with side rails is recommended.)
Patient must be secured with Safety Restraint Strap prior to using the tables.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and 

put into service according to the EMC information provided in the specifications at the end of this manual. 

(Excluding CFPMFXH)
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. (Excluding 

CFPMFXH)
This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous 

oxide.
Electrical Shock Hazard. (Excluding CFPMFXH) The power supply/control module is located inside the lift-

ing column and under the top. No user serviceable parts are inside. Refer servicing to qualified personnel. 

Unplug wall connector prior to contact with any cables connected to the power supply.
To avoid the risk of electric shock, this equipment must only be connected an outlet this is properly 

grounded. See GROUNDING instructions in this manual.
Grounding reliability can only be achieved when this equipment is connected to an equivalent 3 prong 

receptacle marked “Hospital Only” or “Hospital Grade”.
Oakworks medical tables are designed and built to provide many years of dependable service. Please fol-

low local regulations and laws when disposing of the electrical components used in this table.     
Do not apply excessive force to the hand/foot control as it may damage the control.
Do not use the hand/foot control if signs of wear or damage are visible as this can cause unintentional mo-

tion.

Содержание CFPMFXH

Страница 1: ...p CFPMFXH Integrated headrest top Rectangular top CFPM100 Integrated headrest top Rectangular top CFPM300 Rectangular top CFPM301 Integrated headrest top CFPM301 Rectangular top CFPMB301 Rectangular t...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Страница 3: ...le End Extensions 22 Lithotripsy Table used as Urology Table 22 Carbon Fiber Arm Board 23 Adjustable T Rail Clamp 23 Spine Positioning System II 24 Fluoro Extender 24 Reference Usable Imaging Area Dia...

Страница 4: ...orienting the equipment SYMBOL IDENTIFICATION This symbol when used in this manual and on product labels represents a caution warning Be sure to read and comply with all precautions and warnings This...

Страница 5: ...is in motion Be certain that the table is completely lowered prior to discharging an ambulatory patient The patient may lose balance and fall NOTE The CFPMFXH is a fixed height table It is recommended...

Страница 6: ...of 1 10 mm as measured at 100 kVp and a half value HVL of 3 6 mm NOTE Bariatric Top has a typical aluminum filtration equivalence of 1 50 mm as measured at 100 kVp and a half value HVL of 3 6 mm Oakw...

Страница 7: ...ing Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 368 lbs 167 kg OPTIONS 5 13cm Locking Casters 1 2 5cm...

Страница 8: ...sters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 343 lbs 156 kg O...

Страница 9: ...m Locking Casters Locking Unlocking Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all pr...

Страница 10: ...g Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 498 lbs 226...

Страница 11: ...10 cm Locking Casters Locking Unlocking Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during a...

Страница 12: ...40 kg distributed evenly 4 10 cm Locking Casters Locking Unlocking Casters for mobile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint S...

Страница 13: ...ile or permanent positioning Padding 2 5cm Comfort Foam Crescent Face Pad Integrated Headrest Only Patient Restraint Strap Must be used during all procedures Table Weight 548 lbs 249 kg OPTIONS 5 13cm...

Страница 14: ...iolucent dimensions 19 48cm W x 23 7 60cm Table Extensions 1 23 58cm W x 28 71cm L End 1 23 58cm W x 16 41cm L OR 23 58cm W x 20 51cm L Short Table Voltage Options 120V 60Hz or 230V 50Hz 2 8 20cm W x...

Страница 15: ...cm High US Standard Urology Bag Hoop Clips onto T rails bag not included Table Capacity 500 lbs 227 kg distributed evenly 200 lbs 91 kg on all extensions 4 10 cm Locking Casters Locking Unlocking Cast...

Страница 16: ...Clips onto T rails bag not included Table Capacity 500 lbs 227 kg distributed evenly 200 lbs 91 kg on all extensions 4 10 cm Locking Casters Locking Unlocking Casters for mobile or permanent position...

Страница 17: ...oduct is equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropriate outlet that is properly installed and grounded in accordance with...

Страница 18: ...ON CAUTION 15 CAUTION Do not apply excessive force to the foot control as it may damage the control Do not use the foot control if signs of wear or damage are visible as this can cause unintentional m...

Страница 19: ...ateral Tilt OR Vertical Height Trendelenburg Longitudinal Motion Lateral Motion Control functions dependent upon table s fea tures Test each of the table s functions prior to use with a patient CAUTIO...

Страница 20: ...e patient from falling off the table The strap has hook and loop fasteners attached to the ends The strap is attached as follows 1 Place the center of each strap around the patient at two places Reloc...

Страница 21: ...nchored Replace pad by first centering on the table and then pressing flaps in place anchoring with the hook and loop fasteners ADJUSTING THE CRESCENT FACE PADS INTEGRATED HEADREST TOPS ONLY PRONE SUP...

Страница 22: ...the red button To reset rotate the red button clockwise If necessary you can also remove the plug from the outlet Be sure access is not blocked for disconnecting the table from power 19 LOCKING THE CA...

Страница 23: ...e warning line on the tabletop Casters should be locked when patient is mounting or dismounting the table Table should be flat when mounting and dismount ing Patient should mount and dismount near the...

Страница 24: ...Extensions are clamped to the T rail with a locking clamp design There are 2 lateral extensions for the Urology Table 2 5 x 24 13 x 61cm These extensions can be used as required to extend the effectiv...

Страница 25: ...28 58 x 71cm and 1 23 x 16 58 x 41cm Lithotripsy Urology 1 23 x 28 58 x 71cm Urology LITHOTRIPSY TABLE USED AS A UROLOGY TABLE If the extensions are removed from the Lithotripsy Table the table can b...

Страница 26: ...ion of the Arm Board Fold the Base and Support sections for storage Prone Position at 135 Supine Position at 180 Support Section Base Section T RAIL CLAMP USE ACCESSORY 1 The clamp must be installed o...

Страница 27: ...m LARGE RECTANGULAR ADJUSTER PAD 7 X 12 X 1 18 x 31 x 4cm SMALL RECTANGULAR ADJUSTER PAD 8 20cm SEMI ROUND BOLSTER FLUORO EXTENDER ACCESSORY CAUTION Do not place undue weight or downward pressure on t...

Страница 28: ...REFERENCE USABLE IMAGING AREA CFPMFXH CFPM100 INTEGRATED HEADREST TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE 25...

Страница 29: ...26 REFERENCE USABLE IMAGING AREA CFPMFXH CFPM100 RECTANGULAR TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE...

Страница 30: ...REFERENCE USABLE IMAGING AREA CFPM400 CFPM300 CFPM401 CFPM301 INTEGRATED HEADREST TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE 27...

Страница 31: ...28 REFERENCE USABLE IMAGING AREA CFPM400 CFPM300 CFPM401 CFPM301 RECTANGULAR TOPS ONLY CENTERED FOR A P VIEW OBLIQUE 30 VIEW FOR VERTICAL CAUDAL 30 VIEW FOR VERTICAL NO IMAGE NO IMAGE NO IMAGE...

Страница 32: ...LIQUE 30 VIEW FROM VERTICAL CAUDAL 30 VIEW FROM VERTICAL 47 0 119 38cm No Image 23 3 59 06cm 15 5 39 37cm 21 0 53 34cm 2 8 6 99cm TYP 47 0 119 38cm No Image 20 0 50 80cm 19 0 48 26cm 23 0 58 42cm 2 0...

Страница 33: ...IQUE 30 VIEW FROM VERTICAL CAUDAL 30 VIEW FROM VERTICAL 47 0 119 38cm No Image 23 3 59 06cm 15 5 39 37cm 21 0 53 34cm 2 8 6 99cm TYP 47 0 119 38cm No Image 20 0 50 80cm 19 0 48 26cm 23 0 58 42cm 2 0 5...

Страница 34: ...ocedure Because of possible chemical incompatibilities between various cleaning products and bleach utmost care must be taken by the user to avoid potential exposure to harmful or toxic by prod ucts o...

Страница 35: ...nds control and foot control cables and all visible wiring Check the hand foot control for signs of wear or damage replace if necessary Visually inspect components for obvious damage that could cause...

Страница 36: ...window of the Product Label CFPM400 Table Manufacturer Oakworks Inc 923 East Wellspring Road New Freedom PA USA 17349 1 717 235 6807 www oakworks com GTIN SN 00817463021325 QTY 1 EA Manuf date 2016 1...

Страница 37: ...andards 21 CFR Sub chapter J in effect at the time of manufacture for radiographic tables ENVIRONMENTAL CONDITIONS Conditions Temperature Humidity Atmospheric Pressure Normal Use 50 10 C to 104 40 C 2...

Страница 38: ...upporting The table is intended for use in the electromagnetic environment in which radiated RF disturbances are con trolled The customer or the user of the table can help prevent electromagnetic inte...

Страница 39: ...he applicable regulations and requirements of the local state environmental protection agency state health agency Centers for Disease Control CDC Occupational Safety and Health Administration OSHA U S...

Страница 40: ...USER MANUAL FCC Rules and Regulations Title 47 Part 15 Subpart B Class A Unintentional Radiators NOTE This equipment has been tested and found to comply with the limits for a Class A digital device pu...

Отзывы: