8. Plug the new foot control into the control box bracket at the
same connection where the inoperable unit had been.
9. Carefully lift the control box, mounting bracket and excess cords
back up into the cabinet floor, being careful to keep all the excess
out of the way of the control box opening. Be sure all cords are
away from the screw area of the knob connections. Line up the
screw holes, hold the control box bracket base plate in position
and begin to fasten the knobs onto the 4 screws. Hand tighten
them until the bracket is secure.
10. Connect the power cord and test the new foot control.
Should you need assistance at any time, please call a Customer Service
Representative at 800-558-8850.
21
TROUBLESHOOTING
• Foot Control Replacement Instr uctions
T
R
O
U
B
L
E
S
H
O
O
T
IN
G
TR OUBLESHOOTING
• Foot Control