FOOT CONTROL REPLACEMENT INSTRUCTIONS
Replacement of the Foot Control may be necessary if the control does
not actuate the functions of raising or lowering the
C
elesta
G
emini
™
.
Once the replacement unit has been received you are ready to replace
the inoperable foot control.
To Replace the Foot Control, follow these steps:
1. Obtain the
new
Foot Control. (see Parts List, pg. 25)
2. It is recommended that the table be raised to the highest height
level (use hand control) if possible, as you will need to drop the
control box out from beneath the cabinet. You will need about
14" of space to work easily.
3. After raising the table, disconnect the table’s power cord from
your power outlet before proceeding.
4. To access the control box to disconnect the foot control you must first
remove the control box mounting bracket from the underside of
the cabinet (head end) by unscrewing the four knobs that hold the
bracket in place. Brace the bottom of the control box with one hand
while loosening the knobs. Once the knobs are removed, rest the
control box carefully down on the floor. There is plenty of excess cord
stored inside the cabinet to allow the control box to sit on the floor.
19
TROUBLESHOOTING
• Foot Control Replacement Instr uctions
T
R
O
U
B
L
E
S
H
O
O
T
IN
G
TR OUBLESHOOTING
CAUTION
The table weighs as much
as 300 lbs.
DO NOT
lift or
move the unit without using
proper safety procedures
or equipment.
DO NOT
turn
the
C
elesta
G
emini
™
on its side
for any reason. To do so
could cause damage to the
base and
will
void the warranty.
There are no serviceable
items that need this action.
Never lift the table by grasp-
ing the table top as the
table top is not designed to
hold the weight of the base
and you could damage the
table.
IMPORTANT
CAUTION
Before doing any work on the
unit, disconnect the power
cord from the power outlet.
8
• Foot Control
NOTE:
The new electronic
component will be
reinstalled in the same
manner as the old one.
When dissembling please
take note of the order,
installation method and
placement of plugs as
the new unit will be
reassembled in the
same way.
Be sure to read all
cautions, warnings
and instructions given
in the manual to prevent
injur y to both operator
and client.
8
(back of
control box)