![Oakton WD-35610-00 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/oakton/wd-35610-00/wd-35610-00_instructions-for-use-manual_3181834023.webp)
23
11. Specifications
Conductivity
Temperature
Ranges:
0.00 –19.99 µS
0–80°C
0.0 –199.9 µS
(Epoxy platinum probe)
0 –1999 µS
0–100°C
0.00 –19.99 mS
(Glass/Platinum probe)
0.0 –199.9 mS
Resolution:
0.01 µS
0.1 µS
1 µS
0.1°C
0.01 mS
0.1 mS
Conductivity accuracy:
±1% FS in all 3 lower ranges, ±1% FS accuracy within ±20% of calibration
value in the higher ranges. Better accuracies are possible in the very high and very low ranges if
10 and 0.1 cell constant probes (respectively) are used.
Temperature accuracy:
±0.5°C
Cell constant:
1.0 with probe supplied. Cell constant of 0.1 is best for very low ranges (<10 µS)
and a cell constant of 10 is best for higher ranges (100 mS).
Temperature compensation:
auto or manual
Temperature Coefficient Adjustment:
0.0 to 10.0% per °C
Reference temperature:
25°C or 20°C selectable
Calibration:
up to 5 points (one point per range)
Power:
110 or 220 VAC adapter, included
Operating temperature:
0 to 50°C
Dimensions:
meter: 9" x 2
3
⁄
8
" x 7" (229 x 61 x 178 mm)
boxed: 15
1
⁄
2
" x 3
1
⁄
2
" x 10" (394 x 89 x 254 mm)
probe only: 5
1
⁄
2
"L x
1
⁄
2
" dia. (140 mm L x 13 mm dia.) with 2
1
⁄
2
ft (0.8 m) cable
Weight:
meter: 1 lb (0.5 kg)
probe: 0.35 lb (0.2 kg)
warranty/
return of items
appendix
accessories
specifications
error
messages/
troubleshooting
probe care
and
maintenance
setup
program
conductivity
measurement
temperature
calibration
conductivity
calibration
starting up
introduction
Содержание WD-35610-00
Страница 2: ...2...