6. Setup Program
There are two program sets containing several options in
the setup function.
Program 1 (all models):
Stored Values Programs
Program 2 (all models):
General Operation Programs
To Enter Setup
1. Press SET key from the conductivity measurement
(MEAS) mode to enter the SETUP function. The meter
automatically enters Program P1.0. The meter prompts
you with program numbers in the secondary field of
the display.
You must be in the MEAS mode to enter
SETUP.
Using Setup programs
Press the ENTER key to either confirm selection of
options or to move to the next program without changing
the selection. The instrument then automatically scrolls to
the next program. Within each program, you can use
MI/
▲
or MR/
▼
to make selections. Some options only
permit the viewing of data and are useful for diagnostic
purposes.
To exit from the program at anytime after you confirm
your choice or to abort from SETUP, press CAL/MEAS to
return to MEAS. Make sure you press ENTER to confirm
your option in each program.
NOTE:
Exiting the Setup mode without pressing enter
will cause the meter to disregard your new setting
and retain the old setting.
15
Program
Function
Options
Default Setting
P1.0
Memory clear
ON, OFF
OFF
P1.1
% Temperature
Select value
2.00% per °C
Coefficient 0-10%
0-10%/°C
P1.2
Calibration
Indication
-----
status display
only
(not calibrated)
P2.0
Ready On/Off
ON, OFF
ON
P2.1
Auto-Off
ON, OFF
ON
P2.2
Reset
ON, OFF
OFF
P2.3
Temperature
Select
25°C
Normalization
20°C or 25°C
P2.4
Cell constant
Select cell
1.0
selection
constant
Setup programs at a glance
warranty/
return of items
appendix
accessories
specifications
error
messages/
troubleshooting
probe care
and
maintenance
setup
program
conductivity
measurement
temperature
calibration
conductivity
calibration
starting up
introduction