P1.2: Calibration Status Display
P1.2 is a view only program that displays previous calibra-
tion data. This meter can store calibration data for each of
its 5 ranges (one value per range).
NOTE:
The last calibration point entered replaces the
prior calibration point in each range.
1.
Press the MI/
▲
or MR/
▼
keys
to scroll through the
previous calibration data. The measurement range
number (1, 2, 3, 4 or 5) is indicated in the lower left
corner of the display. Ranges that have not been
calibrated will display “----”. The display will also indicate
“µS” or “mS,” depending upon which range was
calibrated.
See Figure 24
.
2.
Press the ENTER key
to move to the next Setup
parameter P 2.0. To exit the SETUP mode, press
CAL/MEAS.
17
Figure 24: P1.2