OaktonLog Plus
Chapter 2
157
3.
Temperature calibration
1.
Click
Logger
on the menu bar, then click
Calibration
.
2.
Enter the calibration password, then click
OK
.
3.
Select
Temperature
or
Ext. Temperature
in the
Choose sensor
drop list.
4.
Click the
Default
button to restore the original values.
5.
Enter the two OaktonLog values into the OaktonLog
Value edit boxes.
6.
Enter the two reference values into the Reference Value
edit boxes.
7.
Click
Calibrate
.
8.
To restore default calibration, click
Default
.