Chapter 2
OaktonLog Plus
154
2.
Humidity calibration
To calibrate the humidity sensor you will need a humidity
chamber.
1.
Connect the OaktonLog to the PC
2.
Set up the OaktonLog to record every 10 seconds (see
page 137).
3.
Click
Logger
on the menu bar, then click
Calibration
.
4.
Enter the calibration password, then click
OK
.
5.
Select
Humidity
in the
Choose sensor
drop list