Wait  for  the  smile  icon  to  stay,  
and then short press      

 

5.3 The tester has already been calibrated after manufacture. Usually users can 

use the tester right away, or test it in the calibration solutions to test its 

accuracy. When error is large, calibrate it before using. 

5.4 We recommend replacing new calibration solutions after 5-10 times of 

calibrations to keep the s

olution’s accuracy.

 

 

6. Parameter Setting 

6.1 Parameter setting reference chart: 

Symbol 

Parameter Setting Items 

Code 

Factory Default 

P1 

TDS Factor 

0.40 to 1.00 

0.71 

P2 

Restore to factory default 

No 

 Yes 

No 

6.2 How to setup parameters: 

When turned off, long press          to enter setup 

 short press           to switch  

P1-P2

 Short press          , parameter flickering

short press            to choose, 

short press          to confirm parameter selection

Long press           to switch off. 

 

6.3 Parameter setting notes 

a) TDS Factor (P1): 

Users can adjust TDS Factor by experimental data or experience. The 

following chart lists some commonly used TDS Factors for reference. 

 

Conductivity and TDS Factor 

Conductivity of the solution 

TDS Factor 

0~100 

μS/cm

 

0.60 

100~1000 

μS/cm

 

0.71 

1~10 mS/cm 

0.81 