  3. Calibration 

3.1 Rinse the probe in distilled water and dry it. Short press            to 

turn on the tester. 

3.2 Long press           to enter calibration mode; Short press          to exit. 

3.3 Dip the probe into 1413 µS calibration solution. Stir gently, leave it to stand, 

LCD displays          (Diagram 3), short press            to complete 1-point calibration. 

The tester returns to measuring mode, and calibration icon          appears at the 

button left of LCD. 

3.4 Rinse probe in distilled water and dry it. Follow the steps in 4.2-4.3 to 

complete 2

nd

 point calibration in 12.88 mS calibration solution, tester returns to 

measuring mode, calibration icons                display on bottom left of LCD. 

 

 

4. TDS Measurement 

4.1 Short press           to turn on tester. Rinse probe in distilled water and dry it. 

4.2 Stir probe in the sample solution gently, leave it to stand. Get readings after 

the smile icon comes up and stays. 

 

5. Notes

 

5.1 The tester adopts 1413 µS and 12.88 mS standard calibration solutions. Users 

can use 1-2 point calibrations as needed. For most circumstances, calibrating in 

1413 µS to complete 1

st

 point calibration will meet testing requirements. 

5.2 The tester has self-diagnosis functions: 

Symbol  Self-Diagnosis information 

How to fix 

 

 

 
 

Wrong calibration solution, 
which exceeds the 
recognizable range of the 
meter. 

1. Check if calibration solution is 
correct 
2. Check if probe is damaged. 

Diagram - 3 

 